FDA Adverse Event Injury Summary report: N

TAPERLOC POR FMRL LAT 10X140

MDR report key: 8817582 · Received July 23, 2019

Report

Report Number
0001825034-2019-03188
Event Type
Injury
Date Received
July 23, 2019
Date of Event
April 22, 2019
Report Date
October 29, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
MEH
PMA / PMN Number
K030055
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. REPORTED EVENT WAS CONFIRMED BY REVIEW OF MEDICAL RECORDS. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. MEDICAL REPORT DETAILS THAT THE PATIENT TURNED WRONG AND HAS INCREASED PAIN WITH SPASMS BILATERAL TOTAL HIP PROSTHESIS¿S APPEAR WELL SEATED. OLD HEALED RIGHT PUBIC RAMUS FRACTURE. NO ACUTE FRACTURE OR DISLOCATION IDENTIFIED. CT REPORT CONFIRMS THERE IS AN AVULSION OF THE LESSER TROCHANTER ON THE POSTERIOR CORTEX OF THE PROXIMAL FEMUR. OLD HEALED PUBIC RAMUS FRACTURE. NO EVIDENCE OF DISLOCATION. METAL LEVELS WERE EXTREMELY HIGH. IMPLANTS THEMSELVES SEEMED STABLE RADIOGRAPHICALLY BACK TO 2012. THERE WAS BLACK STAINING OF THE TISSUE. A COMPLETE SYNOVECTOMY WAS DONE AND METALLOSIS REMOVED. THE ACETABULUM WAS INSPECTED AND THERE WAS NO CLINICAL EVIDENCE OF LOOSENING. THERE IS OSTEOLYSIS OF THE PROXIMAL FEMUR WITH INSUFFICIENCY FRACTURE OF THE LESSER TUBEROSITY FROM THE METALLOSIS AND OSTEOLYSIS. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: M2A 38MM MOD HD -6MM NK # ITEM 11-173660 LOT 934760, M2A 38MMX48MM CUP # ITEM RD118848 LOT 383250. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS:0001825034-2019-02684 AND 0001825034-2019-02685. CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION AS IT REMAINS IMPLANTED. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED

Description of Event or Problem · 1

IT WAS REPORTED, PATIENT TURNED WRONG AND EXPERIENCED SEVERE PAIN AND SPASM. CT CONFIRMED AN AVULSION OF THE LESSER TROCHANTER ON THE POSTERIOR CORTEX OF THE PROXIMAL FEMUR. FURTHER DIAGNOSTICS REVEALED EXTREMELY HIGH SERUM ION LEVELS. THE PATIENT WAS REVISED 16 YEARS POST INITIAL SURGERY FOR METALLOSIS, ELEVATED METAL ION LEVELS, DIFFICULTY AMBULATING, OSTEOLYSIS, BONE FRACTURE, AND PAIN. THERE WAS BLACK STAINING OF THE TISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612750 TAPERLOC POR FMRL LAT 10X140 PROSTHESIS HIP MEH ZIMMER BIOMET, INC. N/A 191410

Patients

Seq Age Sex Outcome Treatment
1 Other