FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 6M M 5.7MM 10MM

MDR report key: 8807155 · Received July 19, 2019

Report

Report Number
0002023141-2019-00446
Event Type
Injury
Date Received
July 19, 2019
Date of Event
June 5, 2018
Report Date
September 23, 2019
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

SIMILAR COMPLAINTS FOR THE INABILITY TO ACHIEVE PRIMARY STABILITY (UNABLE TO PLACE IMPLANT INTO OSTEOTOMY) HAVE BEEN PREVIOUSLY INVESTIGATED. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORDS REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH HAVE NOT SUGGESTED THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. TO DATE, ALL COMPLAINT DATA WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. A REVIEW OF THE SUBJECT LOT (1221996) DID NOT IDENTIFY ANY RELATED NON-CONFORMANCES WHICH WOULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. ADDITIONALLY, THERE HAVE BEEN NO RELATED COMPLAINTS REPORTED AGAINST THE SUBJECT LOT. THESE FACTS DO NOT SUGGEST A SYSTEMIC QUALITY ISSUE WITH THE LOT. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THESE REPORTED EVENTS LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: D4: UNIQUE IDENTIFIER (UDI) NUMBER: (B)(4).

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT (B)(4). INITIAL REPORTER EMAIL ADDRESS: NOT PROVIDED. PMA/510(K) NUMBER: K011028, K013227.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN IMPLANT (TSV6B10) WAS UNABLE TO BE PLACED INTO OSTEOTOMY. IMPLANT DID NOT ACHIEVE PRIMARY STABILITY. SITE WAS BONE GRAFTED. TOOTH LOCATION 28.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602219 IMPL TAPERED SCR-V SBM 6M M 5.7MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL 1221996

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention