PUMP MMT-1715KM 630G 3ML BLACK MEDI
Report
- Report Number
- 2032227-2019-32568
- Event Type
- Injury
- Date Received
- July 19, 2019
- Date of Event
- February 9, 2019
- Report Date
- September 3, 2019
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- UDI-DI
- 00643169873834
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RECEIVED WITH A BLANK DISPLAY DUE TO MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY. UNABLE TO PERFORM THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST, AND THE DAT TEST OR VERIFY KEYPAD UNRESPONSIVE/BUTTON ERROR ALARM DUE TO THE BLANK DISPLAY. MOISTURE DAMAGE WAS ALSO FOUND ON THE MOTOR AND FORCE SENSOR DURING VISUAL INSPECTION.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HYPOGLYCEMIA. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 54 MG/DL. CUSTOMER ALSO STATED THAT SCROLL BAR WAS NOT MOVING WITH NO INPUT AND CENTER, UP AND DOWN ARROWS BUTTONS WAS UNRESPONSIVE. IT WAS ALSO STATED THAT THE BUTTON WAS NOT UNRESPONSIVE DURING AN EASY BOLUS ATTEMPT. CUSTOMER TREATED HYPOGLYCEMIA WITH CEREAL AND YOGURT AND DECLINED TROUBLESHOOTING FOR HYPOGLYCEMIA. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 602401 | PUMP MMT-1715KM 630G 3ML BLACK MEDI | ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1715KM | HG2F94K | 00643169873834 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Other | FRN-UNK-RSVR, UNOMED INF SET |