THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S
Report
- Report Number
- 8010047-2019-02594
- Event Type
- Malfunction
- Date Received
- July 19, 2019
- Date of Event
- June 28, 2019
- Report Date
- September 19, 2019
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- GEI
- UDI-DI
- 04953170409677
- PMA / PMN Number
- K172610
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- NURSE
Narratives
THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. THE TISSUE PAD WAS NOT WORN OR MISSING. THE PROBE WAS BROKEN OFF AT 15 MM FROM THE DISTAL END. THE BROKEN TIP WAS NOT RETURNED. THERE WAS SCRATCH AROUND THE BROKEN POINT. THE FRACTURE SURFACE OF THE PROBE SHOWED THAT CRACK DEVELOPED FROM THE SCRATCH. THE COATING FOR ELECTRIC INSULATION OF THE PROBE AROUND THE SCRATCH WAS PARTIALLY MISSING. THE MANUFACTURING RECORD WAS REVIEWED AND FOUND NO IRREGULARITIES. BASED ON THE PAST SIMILAR CASES, IT WAS KNOWN THAT THE PROBE WAS CRACKED AND THE COATING FOR ELECTRIC INSULATION OF THE PROBE WAS DAMAGED WHEN THE USER ACTIVATED OUTPUT CONTACTING WITH METAL OR SOMETHING HARD OBJECT SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS, BESIDES THE PROBE WAS BROKEN OFF WHEN THE PROBE WAS SUBJECTED TO A LOAD. THE ABOVE DEVICE HANDLING HAS BEEN WARNED IN THE INSTRUCTION MANUAL AS FOLLOWS. DO NOT GRASP OR LET THE PROBE TIP CONTACT HARD OBJECTS SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS (E.G., UTERINE MANIPULATOR). ALSO, BE CAREFUL TO AVOID CONTACTING THE PROBE TIP WITH THOSE ACCIDENTALLY. PARTICULARLY DURING ACTIVATION, A SCRATCH ON THE PROBE TIP COULD OCCUR DUE TO ULTRASONIC VIBRATION, WHICH LEADS THE PROBE TIP TO BREAK AND FALL OFF INTO THE BODY CAVITY. IN ADDITION, THE HIGH-FREQUENCY (RF BIPOLAR) CURRENT FLOWS THROUGH THE METAL AND GENERATES SPARK DISCHARGE, WHICH MAY CAUSE BURNS AND DECREASE FUNCTIONALITIES.
THE SUBJECT DEVICE REFERENCED IN THIS REPORT WAS NOT RETURNED TO OLYMPUS FOR EVALUATION. THEREFORE THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED, IF ADDITIONAL OR SIGNIFICANT INFORMATION BECOMES AVAILABLE AT A LATER TIME.
DURING A LAPAROSCOPIC PROSTATECTOMY, THE SUBJECT DEVICE WAS USED. AFTER 5 MINUTES OF USE, THE TISSUE PAD OF THE SUBJECT DEVICE BROKE OFF INSIDE THE PATIENT. THE FRAGMENT WAS RETRIEVED. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THIS IS THE REPORT REGARDING THE FALLING OF THE TISSUE PAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601941 | THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S | ULTRASONIC SURGICAL DEVICE | GEI | OLYMPUS MEDICAL SYSTEMS CORP. | TB-0535FCS | KR847227 | 04953170409677 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |