FDA Adverse Event Malfunction Summary report: N

SYRINGE 10ML LL TIP BULK CONVENIENCE PAK

MDR report key: 8805472 · Received July 18, 2019

Report

Report Number
1213809-2019-00762
Event Type
Malfunction
Date Received
July 18, 2019
Date of Event
July 1, 2019
Report Date
August 21, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096054
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: ONE PHOTO AND ONE EMPTY TRAY WITHOUT TOP WEB WITH A LOOSE 10ML SYRINGE INSIDE WERE RECEIVED. THE SAMPLE WAS VISUALLY EVALUATED. THE SYRINGE WAS OBSERVED TO HAVE A MOLDING FLASH ON THE PLUNGER ROD¿S THUMB REST. THE FLASH EXTENDED ACROSS THE THUMB REST AND CONTINUED ON TWO OF THE OPPOSING RIBS. THE FLASH OBSERVED WAS OUTSIDE OF THE ALLOWED SPECIFICATION AND IS REJECTABLE. POTENTIAL ROOT CAUSE FOR THE FLASH DEFECT IS ASSOCIATED WITH THE MOLDING PROCESS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT HAS BEEN REPORTED THAT ONE SYRINGE 10ML LL TIP BULK CONVENIENCE PAK HAS BEEN FOUND WITH A MOLDING DEFECT BEFORE USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE IS PARTICULATE FOUND ON THE SYRINGE. PER EMAIL: THERE IS PARTICULATE ON THE TOP OF THE SYRINGE FROM LOT 9002961.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT HAS BEEN REPORTED THAT ONE SYRINGE 10ML LL TIP BULK CONVENIENCE PAK HAS BEEN FOUND WITH A MOLDING DEFECT BEFORE USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE IS PARTICULATE FOUND ON THE SYRINGE. PER EMAIL: THERE IS PARTICULATE ON THE TOP OF THE SYRINGE FROM LOT 9002961.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
594935 SYRINGE 10ML LL TIP BULK CONVENIENCE PAK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 9002961 30382903096054

Patients

Seq Age Sex Outcome Treatment
1 Other