FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK 50ML LUER-LOK SYRINGE

MDR report key: 8803958 · Received July 18, 2019

Report

Report Number
3003152976-2019-00473
Event Type
Malfunction
Date Received
July 18, 2019
Date of Event
June 27, 2019
Report Date
September 4, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
UDI-DI
00382903008650
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 1905217, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 1905217 WERE USED TO CONDUCT A LEAKAGE TEST. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, AND NO LEAK WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A BD PLASTIPAK¿ 50ML LUER-LOK SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "DURING THE PREPARATION AT THE PHARMACY THEY NOTICED SEVERAL TIMES THAT THERE IS A LEAKAGE (CYTOSTATICS) AT THE LEVEL OF THE PLUNGER."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A BD PLASTIPAK¿ 50ML LUER-LOK SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "DURING THE PREPARATION AT THE PHARMACY THEY NOTICED SEVERAL TIMES THAT THERE IS A LEAKAGE (CYTOSTATICS) AT THE LEVEL OF THE PLUNGER."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
599284 BD PLASTIPAK 50ML LUER-LOK SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 1905217 00382903008650

Patients

Seq Age Sex Outcome Treatment
1 Other