FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY

MDR report key: 8800318 · Received July 17, 2019

Report

Report Number
3013756811-2019-41225
Event Type
Malfunction
Date Received
July 17, 2019
Date of Event
June 27, 2019
Report Date
July 17, 2019
Manufacturer
TANDEM DIABETES CARE
Product Code
OZO
UDI-DI
00853052007264
PMA / PMN Number
P180008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT MULTIPLE OCCLUSION ALARMS OCCURRED. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 140-175 MG/DL. A SYSTEM CHECK WAS PERFORMED WITH TANDEM TECHNICAL SUPPORT AND THE OCCLUSION ALARMS WERE VERIFIED. REPORTEDLY, THE CARTRIDGE WAS CHANGED TO ADDRESS THE ISSUE AND INSULIN DELIVERY WAS RESUMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590527 T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO TANDEM DIABETES CARE 1000096 00853052007264

Patients

Seq Age Sex Outcome Treatment
1 56 YR INFUSION SET: AUTOSOFT XC, INSULIN: NOVOLOG