ARCHITECT C16000 SYSTEM
Report
- Report Number
- 1628664-2019-00501
- Event Type
- Malfunction
- Date Received
- July 16, 2019
- Report Date
- August 6, 2019
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- UDI-DI
- 00380740005924
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DURING THE SUBSEQUENT SITE VISIT, THE ABBOTT FIELD SERVICE REPRESENTATIVE (FSR) FOUND THAT THE WASH PRESSURE FOR R1A AND B WERE BELOW SPECIFICATIONS. THE FSR REPLACED THE BELLOWS, BELLOWS ONLY, PART NUMBER 2-89054-02, TO RESOLVE THE ISSUE. AFTER THE R1 WASH PRESSURE WAS DETERMINED TO BE WITHIN SPECIFICATION THE FSR RAN ICT PRECISION AND QC WITH RESULTS WITHIN SPECIFICATIONS. NO ADDITIONAL DISCREPANT RESULTS WERE REPORTED ON THE ARCHITECT SERIAL NUMBER (SN) (B)(6) AFTER THE PART WAS REPLACED. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. A REVIEW OF THE ARCHITECT C1600298 INSTRUMENT SERVICE HISTORY REVEALED NO ADDITIONAL ERRATIC OR DISCREPANT PATIENT RESULTS REPORTED. THERE WERE NO SERVICE OR COMPLAINT ISSUES ON OR AROUND THE DATE THIS COMPLAINT WAS INITIATED THAT MAY HAVE CONTRIBUTED TO THIS ISSUE. THE MONTHLY PRODUCT MONITORING REVIEW FOR THE C16000 SYSTEMS WAS REVIEWED AND NO SYSTEMIC ISSUES OR TRENDS FOR ERRATIC/DISCREPANT RESULTS WAS IDENTIFIED. A REVIEW OF HISTORICAL DATA FOR THE BELLOWS, BELLOWS ONLY (ROHS), PART NUMBER 2-89054-02 FOUND NO TRENDS OR SYSTEMIC ISSUES. THE ARCHITECT SYSTEM OPERATIONS MANUAL AND THE ARCHITECT C16000 SERVICE AND SUPPORT MANUAL, PROVIDE ADEQUATE INFORMATION, TROUBLESHOOTING, AND MAINTENANCE CONCERNING ERRATIC / DISCREPANT RESULTS. INCLUDING, BUT NOT LIMITED TO, THE TROUBLESHOOTING PERFORMED BY THE FIELD SERVICE REPRESENTATIVE (REPLACING THE BELLOWS AND THE POPPET VALVES). BASED ON THE INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED FOR THE ARCHITECT, SN (B)(6) OR THE BELLOWS, BELLOWS ONLY (ROHS) PN 2-89054-02.
ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER REPORTED FALSELY ELEVATED SODIUM (NA) RESULTS ON A PATIENT GENERATED ON THE ARCHITECT ANALYZER. THE RESULTS PROVIDED WERE GENERATED BETWEEN 24JUN2019 THRU 26JUN2019: NA INITIAL WAS 160 AND REPEATED AS NORMAL. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589488 | ARCHITECT C16000 SYSTEM | AUTOMATED CHEMISTRY ANALYZER | JJE | ABBOTT MANUFACTURING INC | 00380740005924 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CONCENTRATED ICT DILUENT, LN 02P32-11| CONCENTRATED ICT DILUENT, LN 02P32-11| LOT # UNKNOWN| LOT # UNKNOWN - SODIUM| CONCENTRATED ICT DILUENT, LN 02P32-11| LOT # UNKNOWN - SODIUM |