FDA Adverse Event Injury Summary report: N

3.5MM LCP PLATE 10 HOLES 137MM

MDR report key: 8793125 · Received July 16, 2019

Report

Report Number
2939274-2019-59194
Event Type
Injury
Date Received
July 16, 2019
Report Date
June 19, 2019
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HRS
UDI-DI
10886982157997
PMA / PMN Number
K082807
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE HISTORY LOT , PART NUMBER: 223.601, LOT NUMBER: H042248, PART MANUFACTURE DATE: 24-FEB-2016, MANUFACTURING LOCATION: ELMIRA, PART EXPIRATION DATE: N/A. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO COMPLAINT RELATED ANOMALIES. THE DEVICE HISTORY RECORD SHOWS THIS LOT OF 3.5MM LCP PLATE 10 HOLES 137MM PRODUCT WAS PROCESSED THROUGH THE NORMAL MANUFACTURING AND INSPECTION OPERATIONS WITH NO REWORK NOR NONCONFORMITIES NOTED. THIS LOT MET ALL DIMENSIONAL, VISUAL AND PACKAGING CRITERIA AT THE TIME OF RELEASE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. A REVIEW OF THE RAW MATERIAL DEVICE HISTORY RECORD REVEALED THIS LOT MET ALL SPECIFICATIONS WITH NO NONCONFORMANCE NOTED. THIS RAW MATERIAL LOT MET ALL DIMENSIONAL AND VISUAL CRITERIA AT THE TIME OF MANUFACTURE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THIS LOT MET ALL DIMENSIONAL, VISUAL AND PACKAGING CRITERIA AT THE TIME OF RELEASE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL PRODUCT CODES HWC, KTT. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. (B)(4). THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT REMOVAL OF PLATES AND SCREWS DUE TO MAL-REDUCTION OF AN ULNA FRACTURE THAT WAS PLATED AT A DIFFERENT LOCATION. THE PATIENT WAS ORIGINALLY IMPLANTED WITH ONE (1) LOCKING COMPRESSION PLATE AND SIX (6) UNKNOWN CORTEX SCREWS. THE SURGEON VERBALIZED THAT IT WAS DONE APPROXIMATELY 7 MONTHS AGO AND MAYBE SCREW PLACEMENT WAS THE REASON FOR FAILURE. ALL HARDWARE ARE INTACT. THERE WAS NO SURGICAL DELAY. PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT OUTCOME IS COMPLETE. THIS IS REPORT 1 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
589375 3.5MM LCP PLATE 10 HOLES 137MM PLATE, FIXATION, BONE HRS WRIGHTS LANE SYNTHES USA PRODUCTS LLC 223.601 H042248 10886982157997

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention