FDA Adverse Event Injury Summary report: N

MINIMAX CEMENTLESS ANATOMICAL STEM RIGHT SIZE 2

MDR report key: 8789471 · Received July 15, 2019

Report

Report Number
3005180920-2019-00613
Event Type
Injury
Date Received
July 15, 2019
Date of Event
July 1, 2019
Report Date
July 15, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030803185
PMA / PMN Number
K170845
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 04 JULY 2019: LOT 169043: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 09-MAY-2017. EXPIRATION DATE: 2022-04-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: PARTIAL HIP REVISION SURGERY (STEM, HEAD AND LINER) PERFORMED 7 MONTHS AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY IN A (B)(6) YEAR OLD WOMAN. ACCORDING TO THE REPORT, THE REASON FOR REVISION IS LEG LENGTH DIFFERENCE DUE TO SEVERE SCOLIOSIS THAT CANNOT BE CONFIRMED ON THE BASIS OF THE PARTIAL X-RAY IMAGE PROVIDED. EXPLANT PICTURES SHOW THE PRESENCE OF BONE ON THE STEM SURFACE SUGGESTING GOOD OSSEOINTEGRATION AND FIXATION OF THE STEM. ON THE BASIS OF INFORMATION AVAILABLE, THERE IS NO REASON TO SUSPECT THAT THIS EVENT IS DUE TO A FAULTY DEVICE. PRELIMINARY INVESTIGATION PERFORMED BY R&D PROJECT MANAGER: PRELIMINARY VISUAL INVESTIGATION SHOWS STEM COVERED IN BIO FLUIDS AND MASSIVE BONE STOCK ADHESION ON THE LATERAL SIDE. ADDITIONAL IMPLANTS INVOLVED, BATCH REVIEW PERFORMED ON 04 JULY 2019. BALL HEADS: MECTACER 01.29.206 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 32 SIZE L +4 (K112115) LOT. 183349. LOT 183349: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18-SEP-2018. EXPIRATION DATE: 2023-09-04. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. LINER: MECTACER 01.29.409 CERAMIC LINER Ø 32 / D LOT. 165467 (DEVICE NOT DISTRIBUTED IN THE USA). LOT 165467: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-DEC-2016. EXPIRATION DATE: 2021-11-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

ON (B)(4) 2019 WE WERE ALERTED OF A REVISION PERFORMED ON (B)(6) 2019. REVISION SURGERY HAS BEEN PERFORMED 6 MONTHS AFTER PRIMARY DUE TO LEG LENGTH DISCREPANCY, 15 MM (APPARENTLY DUE TO SEVERE SCOLIOSIS OF THE SPINE). THROUGH A POSTERIOR APPROACH THE SURGEON REMOVED THE STEM, LINER AND HEAD. HE REPLACED THESE EXPLANTED ITEMS WITH AN MPACT HOODED LINER AND A COMPETITOR REVISION STEM AND HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
581745 MINIMAX CEMENTLESS ANATOMICAL STEM RIGHT SIZE 2 CEMENTLESS HIP STEM LZO MEDACTA INTERNATIONAL SA 169043 07630030803185

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention