FDA Adverse Event Summary report: N

CONTINUOUS WAVE III

MDR report key: 878895 · Received June 21, 2007

Report

Report Number
878895
Date Received
June 21, 2007
Date of Event
May 17, 2007
Report Date
June 21, 2007
Manufacturer
ARTHREX ARTHROSCOPY INSTRUMENTS, INC.
Product Code
HRX
Report Source
User Facility report
Reporter Location
MI, US

Narratives

Description of Event or Problem · 1

DURING A RIGHT KNEE ARTHROSCOPY, THE ARTHREX WATER PUMP BEGAN ALARMING "PRESSURE FAULT." THE CIRCULATING NURSE STOPPED THE PUMP AND BEGAN LOOKING AT THE TROUBLESHOOTING PROCEDURE. THE NURSE LOOKED AT THE FLUID, AND IT WAS ADEQUATE; SHE THEN TOOK THE SUCTION TUBING FROM THE CANISTER OFF, AND TRIED TO RESTART THE PUMP. THE PUMP SPUN VERY FAST AND AGAIN STATED "PRESSURE FAULT." THE NURSE THEN RECONNECTED THE SUCTION AS FLUID WAS COMING OUT OF THE KNEE FROM THE FIELD. ALL THE CLAMPS ALONG THE TUBING WERE CHECKED, AND ALL WERE FOUND TO BE OPEN. THE PUMP WAS RESTARTED, AND AGAIN IT WAS SPINNING REALLY FAST, SO THE PUMP WAS STOPPED AGAIN. AT THIS POINT, THE SURGEON STATED THAT THE PATIENT'S THIGH WAS VERY HARD AND HE WAS HAVING DIFFICULTY BENDING THE KNEE. THE SCOPE WAS REMOVED FROM THE KNEE AND THE TOURNIQUET WAS RELEASED AFTER ONLY 24 MIN OF BEING INFLATED. THE SURGEON THEN APPLIED PRESSURE TO TRY TO REDUCE THE SWELLING. THE NURSE CONTINUED TO TROUBLESHOOT THE PUMP BY REQUESTING A NEW ARTHREX PUMP. THE TUBING WAS CHANGED OVER TO THE NEW MACHINE AND RESTARTED. AFTER ONE MINUTE THE SAME "PRESSURE FAULT" ERROR OCCURRED, HOWEVER, THIS TIME THE SCOPE WAS NOT IN THE KNEE. THE TUBING WAS CHANGED ON THE NEW PUMP, AND THE PUMP WAS RESTARTED. EVERYTHING SEEMED TO CHECK OUT AND WORK CORRECTLY, AND THE CASE WAS COMPLETED. UNFORTUNATELY THE ORIGINAL TUBING USED WAS DISCARDED AFTER THE CASE WAS FINISHED. THIS IS AT LEAST THE THIRD TIME A PROBLEM OF THIS NATURE HAS HAPPENED AT OUR FACILITY. THIS SYSTEM REQUIRES THAT THE EXACT SETUP STEPS BE FOLLOWED PRECISELY IN THE CORRECT ORDER FOR THE PUMP TO FUNCTION PROPERLY. IF THE BALLOON INSIDE THE CYLINDER IS INADVERTENTLY COLLAPSED, PRESSURE CHANGES WILL NOT BE CORRECTLY DETECTED. THE BALLOON MAY BECOME COLLAPSED BY LEAVING THE TUBING CLAMPS ON WHEN STARTING THE PUMP, OR BY SPIKING THE FLUID BAGS BEFORE INSERTING THE DISPOSABLE PRESSURE SENSING COMPONENTS. IF THE TUBING IS DISCONNECTED AND THEN RECONNECTED, OR IF THE SAME TUBING IS USED ON A DIFFERENT MACHINE, THIS WILL NOT REPAIR THE PRESSURE CALIBRATION PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTINUOUS WAVE III PUMP, ARTHROSCOPIC HRX ARTHREX ARTHROSCOPY INSTRUMENTS, INC. AR-6475 *
2 TUBING, ARTHROSCOPIC PUMP --- ARTHREX ARTHROSCOPY INSTRUMENTS, INC. AR-6410 *

Patients

Seq Age Sex Outcome Treatment
1 36 YR