FDA Adverse Event Malfunction Summary report: N

SROM KNEE AUG BLK CEM CLAMP

MDR report key: 8785240 · Received July 12, 2019

Report

Report Number
1818910-2019-98180
Event Type
Malfunction
Date Received
July 12, 2019
Date of Event
January 1, 2019
Report Date
June 21, 2019
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
HXD
UDI-DI
10603295180722
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDED: EVALUATION CODES. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISION OF 21 CFR, PART 803. THE REPORT MAY BE BASED ON THE INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PRODUCT COMPLAINT # (B)(4). EXAMINATION OF THE RETURNED DEVICE CONFIRMED THE REPORTED EVENT. THE NOTED DAMAGE IS CONSISTENT WITH MATERIAL OVERLOAD THROUGH THE USE OF EXCESSIVE TORQUE AND THE INVESTIGATION DID NOT ESTABLISH A NEED FOR CORRECTIVE ACTION. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : NULL. DEVICE HISTORY BATCH : NULL. DEVICE HISTORY REVIEW : NULL. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BOTTOM BLACK RUBBER CLAMP PIECE WAS LOOSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
580009 SROM KNEE AUG BLK CEM CLAMP KNEE INSTRUMENT : CLAMPS HXD DEPUY ORTHOPAEDICS INC US P0494 10603295180722

Patients

Seq Age Sex Outcome Treatment
1