INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2019-09013
- Event Type
- Injury
- Date Received
- July 12, 2019
- Date of Event
- June 5, 2019
- Report Date
- November 12, 2019
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 10888628032910
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: WEIGHT TO THE SPEC, DEFORMATION, BROWN PARTICLES AND CREASE FOLD. CLOUDY WAS OBSERVED AFTER AUTOCLAVE DISINFECTION PROCESS. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO ISSUES FOUND RELATED WITH THE MANUFACTURING PROCESS.
DEVICE WAS EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. THE EVENT OF CAPSULAR CONTRACTURE, BAKER GRADE III IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE III. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 578523 | INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 3249137 | 10888628032910 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Required Intervention |