FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 8783500 · Received July 11, 2019

Report

Report Number
1645337-2019-15233
Event Type
Injury
Date Received
July 11, 2019
Date of Event
June 19, 2019
Report Date
June 19, 2019
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000266
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2019, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 8/5/2019, IT WAS REPORTED TO MENTOR THAT THE REPLACEMENT IMPLANT WAS MENTOR MEMORYGEL BREAST IMPLANT 300CC. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE AND CAPSULAR CONTRACTURE. CONCOMITANT PRODUCTS: MENTOR MEMORYGEL BREAST IMPLANT 250CC, CATALOG 3502501BC, SN (B)(4), LOT 5910838. (B)(4).

Additional Manufacturer Narrative · 1

ON (B)(6) 2019, DURING EVALUATION OF THE SAMPLE IT WAS NOTICED TO BE RUPTURED SINCE WAS RECEIVED IN TWO PARTS. SHELL ABRASION WAS OBSERVED WITHIN THE TEAR. NO OTHER ANOMALIES WERE DISCOVERED. THE SECOND PRODUCT RECEIVED IS RELATED TO A CONCOMITANT DEVICE, THEREFORE NO FURTHER INVESTIGATION IS REQUIRED. THE SHELL ABRASION FAILURE MAY BE THE RESULT OF THE CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE CAUSED BY THE CAPSULAR CONTRACTURE. THE MANUFACTURING RECORDS WERE REVIEWED, AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT. POSTOPERATIVE FORMATION OF A FIBROUS TISSUE CAPSULE AROUND A MAMMARY PROSTHESIS IS A NORMAL PHYSIOLOGIC RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT AND OCCURS IN ALL PATIENTS IN VARYING DEGREES. IN SOME CASES, CONTRACTURE OF THE FIBROUS CAPSULE MAY OCCUR. THIS CONDITION HAS ALSO BEEN ASSOCIATED WITH DEVICE DEFLATION/ RUPTURE, WRINKLING AND/OR DISPLACEMENT OF THE PROSTHESIS. CAPSULAR CONTRACTURE MAY BE MORE COMMON FOLLOWING INFECTION, HEMATOMA, AND SEROMA, AND THE CHANCE OF IT HAPPENING MAY INCREASE OVER TIME. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. BASED ON THE INFORMATION REPORTED AND/OR THE PRODUCT INVESTIGATION, THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH THE MANUFACTURING PROCESS. COMPLAINT INFORMATION WILL BE INCLUDED IN COMPLAINT TRENDING THAT IS REVIEWED AND MONITOR BY QUALITY ASSURANCE ON A REGULAR BASIS TO DETERMINE IF FURTHER ACTION IS NECESSARY. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6)-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A MENTOR MEMORYGEL BREAST IMPLANT 250CC AND EXPERIENCED LEFT IMPLANT RUPTURE AND CAPSULAR CONTRACTURE BAKER GRADE III. AS A RESULT, THE PATIENT WILL UNDERGO EXPLANTATION ON (B)(6) 2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
575732 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 5915685 00081317000266

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention