FDA Adverse Event
Other
Summary report: N
CASTLE
MDR report key: 87808
·
Received April 25, 1997
Report
- Report Number
- 1314329-1997-00002
- Event Type
- Other
- Date Received
- April 25, 1997
- Date of Event
- April 8, 1997
- Report Date
- April 14, 1997
- Manufacturer
- GETINGE/CASTLE, INC.
- Product Code
- FLE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
A SLOWLY LEAKING STEAM-TO-CHAMBER VALVE CAUSED THE CHAMBER TO PARTIALLY FILL WITH HOT WATER, WITH NO INCREASE IN PRESSURE IN THE CHAMBER. A WARNING LIGHT LABELED, "CAUTION DO NOT OPEN DOOR WHEN LIGHT IS ON" WAS ILLUMINATED, INDICATING PRESENCE OF WATER IN CHAMBER. THE OPERATOR IGNORED THE LIGHT AND OPENED THE DOOR. THE RESULTING FLOW OF HOT WATER CAUSED 2ND DEGREE BURNS. THE VALVE WAS DIAGNOSED AND REPAIRED BY A PRODUCT SUPPORT REPRESENTATIVE AND THE UNIT WAS TESTED TO CONFIRM THAT THE MACHINE WAS OPERATING PROPERLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CASTLE | INSTRUMENT WASHER STERILIZER | FLE | GETINGE/CASTLE, INC. | 3025 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Required Intervention |