BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2019-00790
- Event Type
- Malfunction
- Date Received
- July 10, 2019
- Date of Event
- June 21, 2019
- Report Date
- July 16, 2019
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- PMA / PMN Number
- SEE H.10
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR FOREIGN MATTER IN THE GEL WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND FOREIGN MATTER IN THE GEL WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCE'S DURING MANUFACTURING OF THE PRODUCT. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR FOREIGN MATTER IN THE GEL WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS CONDUCTED AND FOREIGN MATTER IN THE GEL WAS OBSERVED. ROOT CAUSE DESCRIPTION: BASED ON THE INVESTIGATION, THE MOST LIKELY ROOT CAUSE WAS DETERMINED TO BE RELATED TO A MANUFACTURING ISSUE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT THERE WAS FOREIGN MATTER IN THE GEL WITH A BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: FM IN THE GEL.
INITIAL REPORTER PHONE #: UNKNOWN. PMA / 510(K)#: K023331, K050036. DEVICE EVALUATED BY MFR: A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THERE WAS FOREIGN MATTER IN THE GEL WITH A BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FM IN THE GEL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 569820 | BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | 8312768 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |