FDA Adverse Event Malfunction Summary report: N

INTIMA-II Y 24GAX0.75IN MZ PRN SLM NPVC

MDR report key: 8773881 · Received July 9, 2019

Report

Report Number
3006948883-2019-00525
Event Type
Malfunction
Date Received
July 9, 2019
Date of Event
May 28, 2019
Report Date
August 1, 2019
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 8233333. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. UNFORTUNATELY A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING. WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A INTIMA-II Y 24GAX0.75IN MZ PRN SLM NPVC. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "THE PATIENT WAS HOSPITALIZED FOR COUGH, NOTICED PRN SLIDING AND RESULT IN DOXOFYLINE LEAKAGE AT PRN DURING INFUSION."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A INTIMA-II Y 24GAX0.75IN MZ PRN SLM NPVC. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "THE PATIENT WAS HOSPITALIZED FOR COUGH, NOTICED PRN SLIDING AND RESULT IN DOXOFYLLINE LEAKAGE AT PRN DURING INFUSION."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
565214 INTIMA-II Y 24GAX0.75IN MZ PRN SLM NPVC INTERVASCULAR CATHETER FOZ BD (SUZHOU) 8233333

Patients

Seq Age Sex Outcome Treatment
1 Other