BD MICROTAINER® MAP MICROTUBE FOR AUTOMATED PROCESS K2 EDTA
Report
- Report Number
- 1917413-2019-01729
- Event Type
- Malfunction
- Date Received
- July 9, 2019
- Date of Event
- June 21, 2019
- Report Date
- September 17, 2019
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JKA
- UDI-DI
- 50382903637060
- PMA / PMN Number
- K093972
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD ACKNOWLEDGES THAT THE CUSTOMER HAS HAD AN ISSUE WITH THE INCIDENT LOT. BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. ADDITIONALLY RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR TESTING. THE SAMPLES WERE EVALUATED AND NO VISIBLE DEFECTS WERE DETECTED, FURTHERMORE PROCESS CAPABILITY DATA FOR THE ADDITIVE DISPENSE EQUIPMENT WERE REVIEWED AND WERE FOUND TO BE WITHIN ACCEPTABLE OPERATIONAL REQUIREMENTS. AS OF 09/11/2019, NO OTHER COMPLAINTS INVOLVING BATCHES 9004697 OR 9042626 HAVE BEEN RECEIVED REPORTING CLOTTING ISSUES. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT WAS CREATED TO CAPTURE THE 4TH OF 6 RELATED INCIDENTS.
IT WAS REPORTED THAT THE SAMPLE FROM A 2-DAY OLD MALE PATIENT COLLECTED IN THE BD MICROTAINER® MAP MICROTUBE FOR AUTOMATED PROCESS K2 EDTA CLOTTED DURING USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 4TH OF 6 RELATED INCIDENTS.
DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE SAMPLE FROM A (B)(6) MALE PATIENT COLLECTED IN THE BD MICROTAINER® MAP MICROTUBE FOR AUTOMATED PROCESS K2 EDTA CLOTTED DURING USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 4TH OF 6 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 5 INCIDENTS ON NICU PATIENTS ON (B)(6) 2019. 1: (B)(6) FEMALE - LOT 9004697 2: (B)(6) MALE - LOT 9042626 3:(B)(6) MALE - LOT 9042626 4: (B)(6) MALE - LOT 9042626 5: (B)(6) MALE - LOT 9004697 6: UNKNOWN DATE, UNKNOWN PT IDENTIFIERS, LOTS 9004697, 9042626 WERE AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 565210 | BD MICROTAINER® MAP MICROTUBE FOR AUTOMATED PROCESS K2 EDTA | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 9042626 | 50382903637060 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2 DA | Other |