BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Report
- Report Number
- 9616657-2019-00251
- Event Type
- Malfunction
- Date Received
- July 9, 2019
- Date of Event
- June 20, 2019
- Report Date
- June 21, 2019
- Manufacturer
- BECTON, DICKINSON AND CO.
- Product Code
- NGT
- UDI-DI
- 30382903065463
- PMA / PMN Number
- SEE H.10.
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE PMA / 510(K)#S REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH 510(K) NUMBER IS AS FOLLOWS: PMA / 510(K)#: K161552. PMA / 510(K)#: K141311. INVESTIGATION SUMMARY: BD WAS UNABLE TO VERIFY OR DETERMINE THE ROOT CAUSE OF THE REPORTED COMPLAINT OF LEAKAGE DUE TO THE RETURNING SAMPLE BEING CONTAMINATED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. INVESTIGATION CONCLUSION: THE SAMPLE RECEIVED WAS CONTAMINATED THEREFORE A COMPLAINT ANALYSIS COULD NOT BE PERFORMED. ROOT CAUSE DESCRIPTION: THE SAMPLE RECEIVED WAS CONTAMINATED THEREFORE A COMPLAINT ANALYSIS COULD NOT BE PERFORMED. AN INVESTIGATION COULD NOT BE COMPLETED, AND A ROOT CAUSE IS UNASSIGNABLE IN THIS CASE. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.
IT WAS REPORTED THAT SALINE LEAKED PAST THE BD POSIFLUSH¿ NORMAL SALINE SYRINGE STOPPER DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NURSE WITNESSED THE SALINE FLUID LEAKING ABOVE PLUNGER AREA OF THE SYRINGE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 564060 | BD POSIFLUSH¿ NORMAL SALINE SYRINGE | SALINE VASCULAR ACCESS FLUSH | NGT | BECTON, DICKINSON AND CO. | 9065683 | 30382903065463 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |