FDA Adverse Event Injury Summary report: N

PNEUMOLINER

MDR report key: 8771649 · Received July 9, 2019

Report

Report Number
9616720-2019-00001
Event Type
Injury
Date Received
July 9, 2019
Date of Event
June 12, 2019
Report Date
July 9, 2019
Manufacturer
ADVANCED SURGICAL CONCEPTS LTD.
Product Code
PMU
PMA / PMN Number
DEN150028
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. A ZERO-DEGREE SCOPE WAS USED FOR VISION INSIDE THE BAG CONTRARY TO THE VALIDATED TRAINING AND IFU WHICH STATES A MINIMUM OF 30 DEGREE SCOPE IS REQUIRED TO GIVE AN ADEQUATE FIELD OF VISION. AN UNTRAINED RESIDENT WAS ACTUATING THE TENACULUM. THE TRAINING AND IFU STIPULATES THAT ONLY A PHYSICIAN WHO HAS COMPLETED THE VALIDATED PNEUMOLINER TRAINING MAY PERFORM THE CONTAINED MORCELLATION PROCEDURE.

Description of Event or Problem · 1

WHILE PERFORMING THE CONTAINED POWER MORCELLATION THE PNEUMOLINER BAG WAS PUNCTURED AND A SECTION OF SMALL BOWEL WAS MORCELLATED. THE PHYSICIAN GRASPED THE PNEUMOLINER BAG WITH THE TENACULUM AND PROCEEDED TO ENGAGE THE POWER MORCELLATOR AND MORCELLATED A SECTION OF THE PNEUMOLINER BAG AND A SECTION OF SMALL BOWEL. THE PATIENT UNDERWENT END TO END ANASTOMOSIS WITH APPROXIMATELY 20 CM OF BOWEL RESECTED. THE PATIENT HAS LEFT HOSPITAL AFTER A RETURN TO NORMAL BOWEL FUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
564760 PNEUMOLINER LAPAROSCOPIC POWER MORCELLATION CONTAINMENT SYSTEM PMU ADVANCED SURGICAL CONCEPTS LTD. WA90500US 113202

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R