ANIMAS VIBE
Report
- Report Number
- 2531779-2019-03905
- Event Type
- Malfunction
- Date Received
- July 8, 2019
- Report Date
- July 3, 2019
- Manufacturer
- ANIMAS LLC
- Product Code
- OYC
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 16-SEP-2019 WITH THE FOLLOWING FINDINGS:. DURING INVESTIGATION, THERE WERE MULTIPLE CALL SERVICE 078-0008 ALARMS IN PUMP HISTORY. DURING REWIND, THE PUMP GIVES A CALL SERVICE 078-0008 ALARM. PUMP OPENED; MOISTURE DAMAGE FOUND ON PCB. ALARM DUE TO MOISTURE DAMAGE. PUMP CASE IS CRACKED NEAR BOTTOM RIGHT CORNER OF DISPLAY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2019, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 078) ISSUE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THIS ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 560409 | ANIMAS VIBE | INSULIN INFUSION PUMP | OYC | ANIMAS LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |