INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC
Report
- Report Number
- 3006948883-2019-00517
- Event Type
- Malfunction
- Date Received
- July 8, 2019
- Date of Event
- June 21, 2019
- Report Date
- July 29, 2019
- Manufacturer
- BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
- Product Code
- FOZ
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9050891& 9050885. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN EITHER PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THESE LOTS WERE ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, A SAMPLE WAS RECEIVED FOR THE PURPOSE OF OUR INVESTIGATION. OUR INVESTIGATORS NOTED A SMALL CRACK WAS FOUND DURING LEAKAGE TESTING THAT WAS LOCATED ON THE ADAPTER. THE RETURNED PRODUCT'S SWAGE DIMENSIONS WERE MEASURED BY OUR TEAM AND FOUND TO BE WITHIN PRODUCT SPECIFICATIONS. HOWEVER ACCORDING TO PREVIOUS INVESTIGATIONS, THROUGH A VARIANCE IN THE AUTOLINE'S SWAGE PRESSURE IT IS POSSIBLE FOR THE MACHINE TO APPLY EXCESS FORCE TO THE DEVICE DURING ASSEMBLY, ALLOWING FOR THE RESULTING CRACK TO FORM IN THE DEVICE. CURRENTLY, WE ARE CONDUCTING A LONG TERM STUDY (CAPA 642738) TO DETERMINE THE ROOT CAUSE FOR THIS EVENT, BUT WE ARE ADDRESSING THE ISSUE THROUGH THE IMPLEMENTATION OF MANUAL INSPECTIONS FOR CRACKS IN THE ADAPTER HEAD, AND WE ARE FURTHUR OPTIMIZING OUR SWAGING PROCESS BY REDUCING DEPTH REQUIREMENTS.
IT WAS REPORTED THAT TWO INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC EXPERIENCED LEAKAGE AT THE CATHETER/HUB JUNCTION DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BLOOD LEAKAGE AT CATHETER HUB WAS NOTICED ONCE COMPLETED PUNCTURE ON PATIENT.
THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9050885, MEDICAL DEVICE EXPIRATION DATE: 03/27/2022, DEVICE MANUFACTURE DATE: 02/19/2019. MEDICAL DEVICE LOT #: 9050891, MEDICAL DEVICE EXPIRATION DATE: 04/05/2022, DEVICE MANUFACTURE DATE: 02/19/2019. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT TWO INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC EXPERIENCED LEAKAGE AT THE CATHETER/HUB JUNCTION DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BLOOD LEAKAGE AT CATHETER HUB WAS NOTICED ONCE COMPLETED PUNCTURE ON PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 559904 | INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC | IV CATHETER | FOZ | BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |