FDA Adverse Event Malfunction Summary report: N

SYNCARDIA COMPANION 2 DRIVER

MDR report key: 8768781 · Received July 8, 2019

Report

Report Number
3003761017-2019-00214
Event Type
Malfunction
Date Received
July 8, 2019
Date of Event
June 23, 2019
Report Date
July 11, 2019
Manufacturer
SYNCARDIA SYSTEMS, LLC
Product Code
LOZ
PMA / PMN Number
P030011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

THE EMERGENCY BATTERY ERROR ALARM AS REPORTED WAS CONFIRMED DURING PATIENT FILE REVIEW. DURING BATTERY EVALUATION A CELL UNDER VOLTAGE (CUV) FLAG AND A PACK UNDER VOLTAGE (PUV) FLAG WERE OBSERVED, HOWEVER NO PERMANENT FAULTS WERE RECORDED. THIS IS AN INDICATION THAT THE EMERGENCY BATTERY WAS DEPLETED AND MAY NEED LONGER THAN THE USUAL ONE HOUR TO FULLY CHARGE. THE EMERGENCY BATTERY WAS ALLOWED TO CHARGE FOR APPROXIMATELY 48 HOURS. AFTER CHARGING, THE BATTERY WAS RE-EVALUATED AND FOUND TO HAVE FULLY CHARGED, THUS CONFIRMING THAT THE BATTERY IS FUNCTIONING AS INTENDED. SINCE THE DRIVER WAS RETURNED WITH THE KEY SWITCH IN THE "ON" POSITION IT IS LIKELY THAT THE DRIVER OPERATED FOR AN EXTENDED PERIOD OF TIME POWERED BY THE EMERGENCY BATTERY, CAUSING IT TO BECOME DEPLETED. THIS ISSUE WILL CONTINUE TO BE MONITORED AND TRENDED AS PART OF THE CUSTOMER EXPERIENCE PROCESS. SYNCARDIA HAS COMPLETED ITS EVALUATION OF THIS COMPLAINT AND IS CLOSING THIS FILE. CE 4864 FOLLOW-UP REPORT 1.

Additional Manufacturer Narrative · 1

THE COMPANION 2 DRIVER IS BEING EVALUATED BY SYNCARDIA. THE RESULTS OF THE INVESTIGATION WILL BE PROVIDED IN A FOLLOW-UP MDR. (B)(4).

Description of Event or Problem · 1

WHILE PERFORMING A ROUTINE EVALUATION, A SYNCARDIA TECHNICIAN REPORTED THAT THE COMPANION 2 DRIVER EXHIBITED AN EMERGENCY BATTERY ERROR ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
562756 SYNCARDIA COMPANION 2 DRIVER EXTERNAL PNEUMATIC DRIVER LOZ SYNCARDIA SYSTEMS, LLC

Patients

Seq Age Sex Outcome Treatment
1