SAFIL UNDYED 4/0 (1,5) 45CM DS19 (M)
Report
- Report Number
- 3003639970-2019-00520
- Event Type
- Malfunction
- Date Received
- July 5, 2019
- Report Date
- July 5, 2019
- Manufacturer
- B.BRAUN SURGICAL SA
- Product Code
- GAM
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, (B)(4) MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K031286. INVESTIGATION: SAMPLES RECEIVED: 39 UNOPENED POUCHES. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED 39 CLOSED SAMPLES FOR ANALYSIS. TIGHTNESS TEST TO THE SAMPLES RECEIVED HAS BEEN PERFORMED AND THE UNITS ARE TIGHT. WE HAVE TESTED THE NEEDLE ATTACHMENT STRENGTH OF ALL SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 1.16 KGF IN AVERAGE AND 0.73 KGF IN MINIMUM (EP REQUIREMENTS: 0.46 KGF IN AVERAGE AND 0.73 KGF IN MINIMUM) . REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. NO CORRECTIVE/PREVENTIVE ACTIONS NEEDED.
IT WAS REPORTED THE THREAD DETACHED FROM THE NEEDLE. THE REPORTER INDICATED THAT THE NEEDLE IS DETACHED FOR THE THREAD, IN ONE CASE WHEN THE PACK WAS OPENED AND IN THREE OTHER INSTANCES THIS OCCURRED DURING A SURGICAL PROCEDURE WHILE THE SURGEON WAS STITCHING. NO PATIENT DATA AVAILABLE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 557566 | SAFIL UNDYED 4/0 (1,5) 45CM DS19 (M) | SYNTHETIC ABSORBABLE BRAIDED S | GAM | B.BRAUN SURGICAL SA | C1049220 | 118206V007 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |