FDA Adverse Event Malfunction Summary report: N

ULTRACISION SURG DEV, BLADES 5MM, ENDO

MDR report key: 876224 · Received September 15, 2006

Report

Report Number
1527736-2006-04675
Event Type
Malfunction
Date Received
September 15, 2006
Date of Event
May 16, 2006
Report Date
June 14, 2006
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
k990362
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY:THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSERT STATES: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENTS ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP CHOLECYSTECTOMY PROCEDURE, HOOKED UP BLADE THROUGH HANDPIECE AND IT WOULD NOT WORK. DID NOT USE INSTRUMENT. GOT A DIFFERENT ONE AND IT WORKED. NO PATIENT CONSQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION SURG DEV, BLADES 5MM, ENDO LFL LFL ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA UNK

Patients

Seq Age Sex Outcome Treatment
1 YR GENERATOR| HANDPIECE