FDA Adverse Event Malfunction Summary report: N

640G INSULIN PUMP MMT-1711K

MDR report key: 8761725 · Received July 4, 2019

Report

Report Number
2032227-2019-24687
Event Type
Malfunction
Date Received
July 4, 2019
Date of Event
April 23, 2019
Report Date
May 19, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
Removal / Correction Number
Z-0958-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0958-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 1

UNIT RECEIVED WITH MISSING RETAINER RING AND RESERVOIR TUBE LIP O-RING. UNIT RECEIVED WITH PARTIALLY BROKEN OFF PIECE AT THE RESERVOIR TUBE LIP. UNABLE TO PERFORM DISPLACEMENT TEST AND P-CAP / RESERVOIR WILL NOT LOCK PROPERLY DUE TO MISSING RETAINER. UNIT RECEIVED WITH CRACKED KEYPAD OVERLAY AT SELECT BUTTON AND MISSING DISPLAY WINDOW COVER. UNIT RECEIVED WITH CRACKED CASE ON THE BACK AT THE BATTERY TUBE AREA, BELT CLIP RAIL CASE CORNER AND BATTERY TUBE THREADS. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS DAMAGED. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 10.3 MMOL/L. CUSTOMER ALSO STATED THAT THERE WAS CRACK ON RESERVOIR RING. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
555492 640G INSULIN PUMP MMT-1711K PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1711K HG18FHN

Patients

Seq Age Sex Outcome Treatment
1 Unknown