FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 8758170 · Received July 3, 2019

Report

Report Number
9617229-2019-07907
Event Type
Injury
Date Received
July 3, 2019
Date of Event
May 1, 2019
Report Date
November 19, 2020
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628007321
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED DEVICE PATCH WITH LOT NUMBER 2903628, CATALOG NUMBER SRF-385, CREASE FOLD, DEFORMATION AND RED PARTICLES ON THE SHELL.

Description of Event or Problem · 0

DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE IV. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
551006 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) SRF-385 2903628 10888628007321

Patients

Seq Age Sex Outcome Treatment
1 28 YR Required Intervention