FDA Adverse Event Injury Summary report: N

ZILVER PTX DRUG-ELUTING PERIPHERAL STENT

MDR report key: 8757357 · Received July 3, 2019

Report

Report Number
3001845648-2019-00301
Event Type
Injury
Date Received
July 3, 2019
Date of Event
October 30, 2017
Report Date
July 3, 2019
Manufacturer
COOK IRELAND LTD
Product Code
NIU
UDI-DI
10827002352756
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

PMA/510(K) #: P100022/S027. COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G. 1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(4). COOK MEDICAL INCORPORATED (CMI) 1025 ACUFF ROAD P.O BOX 4195 BLOOMINGTON INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: 3005580113. DEVICE EVALUATION: THE ZISV6-35-125-5.0-80-PTX DEVICE OF LOT NUMBER C1243651 INVOLVED IN THIS COMPLAINT WAS IMPLANTED IN THE PATIENT, AND WAS NOT AVAILABLE FOR EVALUATION. WITH THE INFORMATION PROVIDED, A DOCUMENT BASED INVESTIGATION WAS CONDUCTED. THERE ARE EIGHT COMPLAINT FILES RELATED TO THIS EVENT. REFER TO COMPLAINT FILES (B)(4): ¿ON CARBON DIOXIDE ANGIOGRAPHY ON (B)(6) 2018, THE DISTAL HALF OF THE LEFT CFA HAD OCCLUDED INTO THE SFA. IMAGING OF THE STUDY STENTS WAS NOT PROVIDED.¿ LAB EVALUATION ¿ N/A. DOCUMENT REVIEW: PRIOR TO DISTRIBUTION ZISV6-35-125-5.0-80-PTX DEVICES ARE SUBJECTED TO VISUAL INSPECTION AND FUNCTIONAL CHECKS TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. A REVIEW OF THE MANUFACTURING RECORDS FOR ZISV6-35-125-5.0-80-PTX OF LOT NUMBER C1243651 DID NOT REVEAL ANY DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT ISSUE. THE REVIEW OF RELEVANT MANUFACTURING RECORDS, CONFIRMS THE FAILURE MODE HAS PREVIOUSLY OCCURRED WITH THE CURRENT LOT NUMBER. THERE IS NO REQUIREMENT TO TAKE ACTION AS RESTENOSIS OF THE STENTED ARTERY IS LISTED AS A KNOWN POTENTIAL ADVERSE EVENT WITHIN THE INSTRUCTIONS FOR USE (IFU) AND PATIENT PRE-EXISTING CONDITIONS CAN CAUSE AND/OR CONTRIBUTE TO RESTENOSIS WITHIN THE STENTED ARTERY. IN THIS CASE THE EVENT OF RESTENOSIS IS RECURRING WITHIN THE SAME PATIENT AND IS MOST LIKELY THE RESULT OF THE PATIENT¿S PRE-EXISTING CONDITIONS OR MEDICAL HISTORY. BASED ON THE INFORMATION AVAILABLE TO DATE, THERE IS NO EVIDENCE TO SUGGEST THAT THERE ARE ANY MANUFACTURING ISSUES ASSOCIATED WITH LOT NUMBER C1243651. THERE IS NO EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE. HOWEVER IT SHOULD BE NOTED THAT RESTENOSIS OF THE STENTED ARTERY IS LISTED AS A KNOWN POTENTIAL ADVERSE EVENT WITHIN THE IFU. IMAGE REVIEW: IMAGES WERE PROVIDED TO SUPPORT THE COMPLAINT INVESTIGATION. THEY WERE REVIEWED THROUGH COOK RESEARCH INC. (CRI) AND THE FOLLOWING COMMENTS WERE PROVIDED BY THE INDEPENDENT REVIEWER: IMPRESSION: 1. THE ADDITIONAL IMAGING COMBINED WITH THE ORIGINALLY PROVIDED IMAGING SUPPORTS A PATTERN OF SFA OCCLUSION FROM ATHEROSCLEROTIC LESIONS PROXIMAL THE STUDY STENTS, RECURRENT SIGNIFICANT INFLOW LIMITATION FROM CFA STENOSIS AND OCCLUSION, AND PROGRESSIVE UNTREATED ILIAC ATHEROSCLEROTIC STENOSIS. ALTHOUGH MODERATE STENOSIS FROM NIH IN PROXIMAL END OF THE ZISV6-35-125-5.0-80-PTX AND THE DISTAL END OF THE ZISV6-35-125-5.0-120-PTX IS CONFIRMED, THE ZISV6-35-125-5.0-80-PTX STENOSIS FORMED IN RESPONSE TO AN ADDITIONAL NON-STUDY STENT AND THE ZISV6-35-125-5.0-120-PTX STENOSIS WAS AGGRAVATED BY INFLOW LIMITATION. 2. THE ZISV6-35-125-5.0-80-PTX NIH WAS A SECONDARY RESPONSE TO INSUFFICIENT OVERLAP WITH A THIRD STENT IMPLANTED JUST SUPERIOR TO THE ZISV6-35-125-5.0-80-PTX. BECAUSE A THIRD LARGER DIAMETER STENT WAS IMPLANTED FLUSH WITH THE ZISV6-35-125-5.0-80-PTX SUPERIOR END RATHER THAN OVERLAPPED INTO THE ZISV6-35-125-5.0-80-PTX, THE SFA AT THIS LOCATION WOULD HAVE BEEN IRRITATED BY THE DIFFERING STENT DIAMETERS, THE ENDS OF THE STENTS, AND UNSUPPORTED BENDING AND TWISTING. APART FROM SMOOTH AS OPPOSED TO SHARP STENT ENDS, THE LACK OF OVERLAP ESSENTIALLY FUNCTIONED AS A STENT FRACTURE. THIS IRRITATION WOULD HAVE PROVOKED NIH. 3. THE ZISV6-35-125-5. 0-120-PTX STENOSIS FOLLOWED A MAJOR UNREPORTED UPSTREAM SFA EVENT, A NEW 65% STENOSIS OF THE LEFT CFA, AND CONTINUED TOBACCO ABUSE. THE ADDITIONAL IMAGING DOCUMENTS INTERVENTIONS AFTER THE COMPLAINT EVENT, NOT BEFORE. THESE EVENTS FOLLOW AT A PATTERN OF OCCLUSION BEGINNING PROXIMAL THE STENTS FOLLOWED BY SHORT TERM, LESS THAN SIX MONTHS, ANGIOGRAPHIC FOLLOW-UP. WHEN THE MODERATE ZISV6-35-125-5.0-120-PTX STENOSIS FROM NIH THAT GENERATED THE COMPLAINT WAS DISCOVERED, IT WAS LIKELY DURING SHORT TERM FOLLOW UP OF THE UNDOCUMENTED EVENT. BECAUSE THE SECOND NON-STUDY STENT WAS IMPLANTED ONLY AFTER FAILED ATHERECTOMY AND ANGIOPLASTY OF A NEW PROXIMAL OCCLUSION, THE FIRST NON-STUDY STENT WAS LIKELY THAT IMPLANTED FOR A SIMILAR LESION. CONSEQUENTLY, THE COMPLAINT STENOSIS INVOLVING THE ZISV6-35-125-5.0-120-PTX WAS LIKELY DISCOVERED INCIDENTAL TO A PROXIMAL OCCLUSION/SEVERE STENOSIS REQUIRING STENT PLACEMENT. 4. THE RECURRENT ZISV6-35-125-5.0-120-PTX STENOSIS, THAT IS NOT THE SUBJECT OF A COMPLAINT, BEGAN ONE YEAR AFTER IMPLANTATION, AFTER ATHERECTOMY, AND WAS DOWNSTREAM RECURRENT SEVERE LEFT CFA STENOSIS. BECAUSE THE STENT WAS ALREADY COVERED WITH INTIMA, THE STENT COULD NOT HAVE HAD A BIOLOGIC EFFECT ON THIS RESTENOSIS. BECAUSE IT WAS LOCATED AT THE ADDUCTOR CANAL WHICH ALONE IS PRONE TO STENOSIS, THE STENT COULD HAVE CONTRIBUTED TO RESTENOSIS ON A MECHANICAL BASIS. 5. RECURRENT RIGHT LEG ATHEROSCLEROTIC STENOSIS AND OCCLUSION WAS INDICATIVE OF SEVERE ATHEROSCLEROTIC DISEASE AND CONSISTENT WITH THE REPORTED CONTINUED TOBACCO ABUSE. ROOT CAUSE REVIEW: A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. A POSSIBLE ROOT CAUSE COULD BE ATTRIBUTED TO PATIENT PRE-EXISTING CONDITIONS. FROM THE INFORMATION PROVIDED IT IS KNOWN THAT THE PATIENT HAD A MEDICAL HISTORY OF HYPERTENSION, HYPERCHOLESTEROLEMIA AND IS A CURRENT SMOKER. IT IS KNOWN FROM THE IMAGING REVIEW THAT THE OCCLUSIONS FORMED AS A RESULT OF RECURRENT SIGNIFICANT INFLOW LIMITATION FROM CFA (COMMON FEMORAL ARTERY) STENOSIS AND OCCLUSION, AND PROGRESSIVE UNTREATED ILIAC ATHEROSCLEROTIC STENOSIS. IT IS POSSIBLE THAT THE PATIENT¿S PRE-EXISTING CONDITIONS CAUSED AND/OR CONTRIBUTED TO THIS EVENT. FROM THE INFORMATION PROVIDED IT IS KNOWN THAT THE SITE INDICATED THAT MULTIPLE PREVIOUS HOSPITALISATIONS FOR THE LOWER LIMBS MAY HAVE CAUSED AND/OR CONTRIBUTED TO THIS EVENT. IT IS KNOWN FROM THE INFORMATION PROVIDED THAT THE STENT DID NOT MALFUNCTION OR DETERIORATE IN CHARACTERISTIC OR PERFORMANCE. SUMMARY: THE COMPLAINT IS CONFIRMED BASED ON CUSTOMER TESTIMONY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT UNDERWENT LEFT ILIO-FEMORAL DEEP THROMBOENDARTERECTOMY, LEFT STENT REMOVAL DUE TO CONDITION, REMOVAL OF A SUPERFICIAL FEMORAL STENT, THROMBECTOMY OF THE LEFT SUPERFICIAL FEMORAL ARTERY, WIDENING PATCH OF THE DEEP MEDIO-CURUAL AND ILIAC ANGIOPLASTY AS A RESULT OF THIS EVENT.

Additional Manufacturer Narrative · 0

PMA/510(K) #: P100022/S027. COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G.1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(4) COOK MEDICAL INCORPORATED (CMI) 1025 ACUFF ROAD P.O BOX 4195 BLOOMINGTON INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: 3005580113. INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.

Description of Event or Problem · 0

BRIEF DESCRIPTION: REASON FOR TRACKWISE ENTRY ¿ OCCLUSION/STENOSIS IN THE STUDY LEG. ON (B)(6) 2017 DURING THE INDEX PROCEDURE THE PATIENT RECEIVED TWO ZILVER© PTX© STENTS. THE STUDY LESION WAS IN THE LEFT PROXIMAL AND DISTAL SFA. THE LESION STATUS WAS DE NOVO AND THERE WAS NO CALCIFICATION OR THROMBUS OBSERVED PRE-PROCEDURE. THE INFLOW TRACT WAS PATENT AND THERE WAS ONE RUNOFF VESSEL. THE LESION LENGTH WAS 200 MM AND THE PROXIMAL RVD WAS 5.10 MM AND DISTAL RVD WAS 5 MM. THE PERCENTAGE DIAMETER STENOSIS IN STUDY LESION WAS 3 %. THE POST-PROCEDURE IMAGING SHOWED NO DIAMETER STENOSIS LEFT IN THE STUDY LESION. THIS WAS QUERIED AND UPDATED FROM 80% TO NONE. ON (B)(6) 2017 ¿ (B)(6) 2017 (80 DAYS POST-PROCEDURE), THE PATIENT WAS HOSPITALIZED FOR AN OCCLUSION OF THE RIGHT (NON-STUDY LEG) FEMORAL ARTERY. A ILIO-FEMORAL BYPASS WAS PERFORMED. THIS WAS NOT RELATED TO DEVICE OR PROCEDURE. ON (B)(6) 2017 ¿ (B)(6) 2017 (184 DAYS POST-PROCEDURE), A STENOSIS OF THE RIGHT FEMORAL ARTERY AND FEMOROPOPLITEAL BYPASS WAS REVEALED. THIS WAS TREATED WITH PLACEMENT OF A STENT. ON (B)(6) 2017 (194 DAYS POST-PROCEDURE), A FOLLOW-UP VISIT REVEALED A STUDY LEG RUTHERFORD CLASSIFICATION OF 3. THE STUDY LESION WAS OCCLUDED. ON (B)(6) 2017 ¿ (B)(6) 2017 (298 DAYS POST-PROCEDURE), AN OCCLUSION OF THE LEFT SFA WAS REVEALED. ANGIOGRAPHY REVEALED 85 % STENOSIS. THIS WAS TREATED WITH AN ENDOVASCULAR PROCEDURE WHERE A STENT WAS PLACED. NO DEVICE MALFUNCTION WAS NOTED BUT THE EVENT WAS THOUGHT TO BE DEFINITELY RELATED TO THE DEVICE. THE RELATIONSHIP TO THE PROCEDURE WAS UNKNOWN. NO PRE-EXISTING CONDITION CAUSED THIS. ON (B)(6) 2018 (368 DAYS POST-PROCEDURE), A FOLLOW-UP VISIT REVEALED A STUDY LEG RUTHERFORD CLASSIFICATION OF 0. THE STUDY LESION WAS PATENT. ON (B)(6) 2017 DURING THE INDEX PROCEDURE THE PATIENT RECEIVED TWO ZILVER© PTX© STENTS. THE STUDY LESION WAS IN THE LEFT PROXIMAL AND DISTAL SFA, 200 MM IN LENGTH. THE LESION STATUS WAS DE NOVO WITH NO CALCIFICATION, NO THROMBUS, A PATENT INFLOW TRACT, AND ONE PATENT RUNOFF VESSEL. THE PROXIMAL REFERENCE VESSEL DIAMETER (RVD) WAS 5.10 MM, THE DISTAL RVD WAS 5.0 MM WITH 100% DIAMETER STENOSIS IN STUDY LESION. THE LESION WAS PRE-DILATED WITH A BARE BALLOON ANGIOPLASTY WITH 80% DIAMETER STENOSIS IN STUDY LESION AFTER PRE-TREATMENT. THERE WERE NO DIFFICULTIES DURING THE PROCEDURE AND THERE WAS NO RESIDUAL STENOSIS AT THE COMPLETION OF THE PROCEDURE. THE PATIENT WAS DISCHARGED ON (B)(6) 2017. ON (B)(6) 2018, THE PATIENT WAS DIAGNOSED WITH AN OCCLUSION/RESTENOSIS WITHIN THE STUDY LESION. THE CLINICAL SIGN AND SYMPTOMS WERE PERSISTENT CLAUDICATION. THE IMAGING EVIDENCE WAS PER ANGIOGRAPHY WHICH REVEALED 100% DIAMETER STENOSIS. THE TREATMENT INCLUDED SURGICAL INTERVENTION, ¿LEFT ILIO-FEMORAL DEEP THROMBOENDARTERECTOMY. LEFT STENT REMOVAL DUE TO CONDITION REMOVAL OF A SUPERFICIAL FEMORAL STENT THROMBECTOMY OF THE LEFT SUPERFICIAL FEMORAL ARTERY WIDENING PATCH OF THE DEEP MEDIO-CRURAL ILIAC ANGIOPLASTY, LEFT PRIMITIVE AND EXTERNAL ARTERIOGRAPHY OF THE RIGHT LOWER LIMB IN CROSS OVER.¿ THE SITE INDICATED THAT THE PRE-EXISTING CONDITION OF ¿MULTIPLE HOSPITALIZATION FOR THE LOWER LIMBS¿ CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE DID NOT MALFUNCTION OR DETERIORATE IN CHARACTERISTICS OR PERFORMANCE. THE SITE WAS QUERIED TO DETERMINE WHICH STENT WAS REMOVED AND INDICATED IT WAS NOT POSSIBLE TO DETERMINE, THE SOURCE DOCUMENTS DID NOT INDICATE WHICH STENT WAS REMOVED. IT IS UNKNOWN IF IT WAS EVEN A STUDY STENT REMOVED "OCCLUSION AND INTERVENTION IN (B)(6) 2018.

Description of Event or Problem · 0

BRIEF DESCRIPTION: REASON FOR (B)(4) ENTRY ¿ OCCLUSION/STENOSIS IN THE STUDY LEG. ON (B)(6) 2017 DURING THE INDEX PROCEDURE THE PATIENT RECEIVED TWO ZILVER© PTX© STENTS. THE STUDY LESION WAS IN THE LEFT PROXIMAL AND DISTAL SFA. THE LESION STATUS WAS DE NOVO AND THERE WAS NO CALCIFICATION OR THROMBUS OBSERVED PRE-PROCEDURE. THE INFLOW TRACT WAS PATENT AND THERE WAS ONE RUNOFF VESSEL. THE LESION LENGTH WAS 200 MM AND THE PROXIMAL RVD WAS 5.10 MM AND DISTAL RVD WAS 5 MM. THE PERCENTAGE DIAMETER STENOSIS IN STUDY LESION WAS 3 %. THE POST-PROCEDURE IMAGING SHOWED NO DIAMETER STENOSIS LEFT IN THE STUDY LESION. THIS WAS QUERIED AND UPDATED FROM 80% TO NONE. ON (B)(6) 2017 ¿ (B)(6) 2017 (80 DAYS POST-PROCEDURE), THE PATIENT WAS HOSPITALIZED FOR AN OCCLUSION OF THE RIGHT (NON-STUDY LEG) FEMORAL ARTERY. A ILIO-FEMORAL BYPASS WAS PERFORMED. THIS WAS NOT RELATED TO DEVICE OR PROCEDURE. ON (B)(6) 2017 (184 DAYS POST-PROCEDURE), A STENOSIS OF THE RIGHT FEMORAL ARTERY AND FEMOROPOPLITEAL BYPASS WAS REVEALED. THIS WAS TREATED WITH PLACEMENT OF A STENT. ON (B)(6) 2017 (194 DAYS POST-PROCEDURE), A FOLLOW-UP VISIT REVEALED A STUDY LEG RUTHERFORD CLASSIFICATION OF 3. THE STUDY LESION WAS OCCLUDED. ON (B)(6) 2017 ¿ (B)(6) 2017 (298 DAYS POST-PROCEDURE), AN OCCLUSION OF THE LEFT SFA WAS REVEALED. ANGIOGRAPHY REVEALED 85 % STENOSIS. THIS WAS TREATED WITH AN ENDOVASCULAR PROCEDURE WHERE A STENT WAS PLACED. NO DEVICE MALFUNCTION WAS NOTED BUT THE EVENT WAS THOUGHT TO BE DEFINITELY RELATED TO THE DEVICE. THE RELATIONSHIP TO THE PROCEDURE WAS UNKNOWN. NO PRE-EXISTING CONDITION CAUSED THIS. ON (B)(6) 2018 (368 DAYS POST-PROCEDURE), A FOLLOW-UP VISIT REVEALED A STUDY LEG RUTHERFORD CLASSIFICATION OF 0. THE STUDY LESION WAS PATENT. ON (B)(6) 2017 DURING THE INDEX PROCEDURE THE PATIENT RECEIVED TWO ZILVER© PTX© STENTS. THE STUDY LESION WAS IN THE LEFT PROXIMAL AND DISTAL SFA, 200 MM IN LENGTH. THE LESION STATUS WAS DE NOVO WITH NO CALCIFICATION, NO THROMBUS, A PATENT INFLOW TRACT, AND ONE PATENT RUNOFF VESSEL. THE PROXIMAL REFERENCE VESSEL DIAMETER (RVD) WAS 5.10 MM, THE DISTAL RVD WAS 5.0 MM WITH 100% DIAMETER STENOSIS IN STUDY LESION. THE LESION WAS PRE-DILATED WITH A BARE BALLOON ANGIOPLASTY WITH 80% DIAMETER STENOSIS IN STUDY LESION AFTER PRE-TREATMENT. THERE WERE NO DIFFICULTIES DURING THE PROCEDURE AND THERE WAS NO RESIDUAL STENOSIS AT THE COMPLETION OF THE PROCEDURE. THE PATIENT WAS DISCHARGED ON (B)(6) 2017. ON (B)(6) 2018, THE PATIENT WAS DIAGNOSED WITH AN OCCLUSION/RESTENOSIS WITHIN THE STUDY LESION. THE CLINICAL SIGN AND SYMPTOMS WERE PERSISTENT CLAUDICATION. THE IMAGING EVIDENCE WAS PER ANGIOGRAPHY WHICH REVEALED 100% DIAMETER STENOSIS. THE TREATMENT INCLUDED SURGICAL INTERVENTION, ¿LEFT ILIO-FEMORAL DEEP THROMBOENDARTERECTOMY. LEFT STENT REMOVAL DUE TO CONDITION REMOVAL OF A SUPERFICIAL FEMORAL STENT THROMBECTOMY OF THE LEFT SUPERFICIAL FEMORAL ARTERY WIDENING PATCH OF THE DEEP MEDIO-CRURAL ILIAC ANGIOPLASTY, LEFT PRIMITIVE AND EXTERNAL ARTERIOGRAPHY OF THE RIGHT LOWER LIMB IN CROSS OVER¿ THE SITE INDICATED THAT THE PRE-EXISTING CONDITION OF ¿MULTIPLE HOSPITALIZATION FOR THE LOWER LIMBS¿ CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE DID NOT MALFUNCTION OR DETERIORATE IN CHARACTERISTICS OR PERFORMANCE. THE SITE WAS QUERIED TO DETERMINE WHICH STENT WAS REMOVED AND INDICATED IT WAS NOT POSSIBLE TO DETERMINE, THE SOURCE DOCUMENTS DID NOT INDICATE WHICH STENT WAS REMOVED. IT IS UNKNOWN IF IT WAS EVEN A STUDY STENT REMOVED. OCCLUSION AND INTERVENTION IN (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
554210 ZILVER PTX DRUG-ELUTING PERIPHERAL STENT NIU STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING NIU COOK IRELAND LTD C1243651 10827002352756

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention