FDA Adverse Event Injury Summary report: N

BERCHTOLD OPERON D860

MDR report key: 8749802 · Received July 1, 2019

Report

Report Number
MW5087728
Event Type
Injury
Date Received
July 1, 2019
Date of Event
May 29, 2019
Report Date
June 19, 2019
Manufacturer
STRYKER COMMUNICATIONS INC.
Product Code
GDC
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
WA, US
Reporter Occupation
003

Narratives

Description of Event or Problem · 1

WHEN THE LEG SPARS ON THE TABLE ARE NOT PRECISELY ALIGNED, PRESSING THE UP OR DOWN LEG BUTTON ON THE HAND PENDANT CAUSES THE LEG SPARS TO CONTINUE MOVING UP OR DOWN ON THEIR OWN AFTER THE BUTTON IS RELEASED. THIS INCLUDES A FULL UPWARD FLEXION OF THE LEG SPARS WHEN THE CABLE IS SUPINE. THE LEG SPARS WILL CONTINUE MOVING THROUGH THEIR FULL RANGE OF MOTION AND CAN ONLY BE STOPPED BY TURNING OFF POWER TO THE TABLE. THESE UNINTENDED MOVEMENTS PLACE PTS AT RISK FOR SERIOUS INJURY. ACCORDINGLY, ALL SIX TABLES HAVE BEEN TAKEN OUT OF USE. REF MW5087729, MW5087730, MW5087731, MW5087732, MW5087733.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
542477 BERCHTOLD OPERON D860 TABLE, OPERATING ROOM, ELECTRICAL GDC STRYKER COMMUNICATIONS INC. D 860

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention