FDA Adverse Event
Malfunction
Summary report: N
MTS PIPETTE
MDR report key: 874853
·
Received July 5, 2007
Report
- Report Number
- 9681721-2007-00010
- Event Type
- Malfunction
- Date Received
- July 5, 2007
- Date of Event
- May 11, 2007
- Report Date
- July 5, 2007
- Manufacturer
- BIOHIT OYJ
- Product Code
- JTC
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER INDICATED THAT, THE BIOHIT PIPETTOR DISPENSED THE INCORRECT AMOUNT OF FLUID. NO DEFINITIVE ROOT CAUSE WAS DETERMINED. THE CUSTOMER OBSERVED THE ISSUE WHILE PIPETTING SAMPLE AND ABORTED TENTING, PREVENTING ERRONEOUS TEST RESULTS FROM BEING REPORTED. HOWEVER, IF THIS INCIDENT WERE TO OCCUR UNDETECTED, AN INCORRECT DISPENSE OF FLUID COULD LEAD TO VARIATION IN ANTIGEN/ANTIBODY RATIO AND ERRONEOUS TEST RESULTS. THE CUSTOMER COMPLAINT WAS CONFIRMED. ISSUE RESOLVED THROUGH PRODUCT REPLACEMENT. THE PIPETTOR IN QUESTION WILL BE STAGED FOR FUTURE DISPOSITION.
Description of Event or Problem · 1
CUSTOMER REPORTED, THE PIPETTE AS MAKING A FUNNY NOISE AND DOES NOT DISPENSE THE CORRECT VOLUME. (CUSTOMER NOTICED PROPER AMOUNT NOT DISPENSE USING PIPETTOR).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MTS PIPETTE | ELECTRONIC PIPETTOR | JTC | BIOHIT OYJ | 710142MT | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |