BD PLASTIPAK 50ML CONCENTRIC LUER LOCK SYRINGE
Report
- Report Number
- 3003152976-2019-00430
- Event Type
- Malfunction
- Date Received
- June 28, 2019
- Date of Event
- June 11, 2019
- Report Date
- August 8, 2019
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- UDI-DI
- 00382903008650
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION SUMMARY: NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 1904223, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. TEN RETAINED SAMPLES OF LOT 1904223 WERE USED TO CONDUCT A LEAKAGE TEST. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, AND NO LEAK WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS. H3 OTHER TEXT : SEE SECTION H.10.
IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "CUSTOMER REPORTED LEAKAGE PAST STOPPER - LOT 1904223".
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT LEAKAGE OCCURRED DURING USE WITH A BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "CUSTOMER REPORTED LEAKAGE PAST STOPPER - LOT 1904223."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 536552 | BD PLASTIPAK 50ML CONCENTRIC LUER LOCK SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 1904223 | 00382903008650 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |