FDA Adverse Event Malfunction Summary report: N

BD PHASEAL¿ INFUSION SET (C50)

MDR report key: 8744128 · Received June 28, 2019

Report

Report Number
3003152976-2019-00429
Event Type
Malfunction
Date Received
June 28, 2019
Date of Event
June 12, 2019
Report Date
August 28, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
LHI
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE CORRECTION IS AS FOLLOWS: MEDICAL DEVICE MANUFACTURER: BECTON DICKINSON, S.A.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON REVIEWING THE PHOTO, IT WAS VERIFIED THE CONNECTOR WAS NOT ATTACHED TO THE DRIP CHAMBER. A DEVICE HISTORY REVIEW WAS PERFORMED AND FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF LOT K71646-1. THE FINAL PRODUCT FOR LOT K71646-1 IS ASSEMBLED AND PACKAGED AT A SUPPLIER SITE. WE NOTIFIED THE SUPPLIER OF THE REPORTED ISSUE AND PROVIDED THE PHOTO FOR EVALUATION. THE ASSEMBLY OF THE CONNECTOR TO THE CHAMBER IS A MANUAL PROCESS WHICH IS COMPLETED PRIOR TO ASSEMBLING THE TUBING. THROUGH THEIR INVESTIGATION IT WAS DETERMINED THAT THIS ISSUE LIKELY OCCURRED AS HUMAN ERROR. THE OPERATOR LIKELY OVERLOOKED THIS TASK AND DID NOT PROPERLY IDENTIFY THE MISSING CONNECTOR BEFORE THE ITEM WAS PACKAGED. IT IS BELIEVED THIS IS AN ISOLATED INCIDENT THAT IS UNLIKELY TO REOCCUR. COMPLAINTS RECEIVED FOR THIS DEVICE AND DEFECT WILL BE CONTINUE TO BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS. CONCLUSION(S): THE FINAL PRODUCT SECONDARY SET C50 IS NOT ASSEMBLED/PACKAGED IN SAN AGUSTÍN FACILITIES. THE INFORMATION OF THIS COMPLAINT WAS SENT TO THE SUBCONTRACTOR (CODAN) TO PERFORM THE INVESTIGATION (SC-058-19). ACCORDING TO THEIR REPORT, THE DEFECT SEEMS TO BE CAUSED BY A HUMAN ERROR. THE ASSEMBLY OF THE DRIP CHAMBER AND CARMEL ADAPTER WAS A MANUAL PROCESS. APPARENTLY, OPERATOR OVERLOOKED THE ASSEMBLY OF THE ADAPTER DUE TO A HUMAN ERROR. THE FAULT IS CONSIDERED AS AN ISOLATED CASE. H3 OTHER TEXT : SEE SECTION H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS A MISSING CONNECTOR WITH A BD PHASEAL¿ INFUSION SET (C50) BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "CONNECTOR PART WAS NOT IN THE PACKAGING."

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS A MISSING CONNECTOR WITH A BD PHASEAL¿ INFUSION SET (C50) BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "CONNECTOR PART WAS NOT IN THE PACKAGING."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A MISSING CONNECTOR WITH A BD PHASEAL¿ INFUSION SET (C50) BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "CONNECTOR PART WAS NOT IN THE PACKAGING."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
536526 BD PHASEAL¿ INFUSION SET (C50) INFUSION SET LHI BECTON DICKINSON, S.A. K71646-1

Patients

Seq Age Sex Outcome Treatment
1 Other