FDA Adverse Event Injury Summary report: N

EPIC PLATE SYSTEM

MDR report key: 8743954 · Received June 28, 2019

Report

Report Number
1038671-2019-05045
Event Type
Injury
Date Received
June 28, 2019
Date of Event
November 7, 2016
Report Date
June 28, 2019
Manufacturer
EPIC EXTREMITY
Product Code
HRS
PMA / PMN Number
K153340
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT DEVICES WERE NOT RETURNED FOR ANALYSIS. A PULL THROUGH TEST WAS CONDUCTED USING SIMILAR PARTS. THE SCREWS WERE TIGHTENED AGAINST THE PLATE ON-AXIS (+/- 12.5-DGREES), AT 20-DEGREES OFF-AXIS, AND 30-DEGREES OFF-AXIS. WHEN USING A 5MM DIAMETER HEAD (ORIGINAL HEAD DESIGN), STAYING WITHIN +/-12.5 DEGREE OFF AXIS ALLOWS FOR PROPER INSTALLATION. WHEN OFF AXIS IS GREATER THAN 12.5 DEGREES RISK IS INTRODUCED TO PULL SCREW THROUGH PLATE, LIKELIHOOD OF RISK INCREASES AS OFF AXIS ANGLE INCREASES. FURTHER TESTING WAS CONDUCTED USING THE NEWLY DESIGNED LOCKING SCREWS. WHEN USING 5.5MM DIAMETER HEAD (NEW HEAD DESIGN), PROPER INSTALLATION WAS ACHIEVED AT ALL OFF AXIS ANGLES UP TO +/- 30 DEGREES OFF AXIS, WHICH WAS THE SCREW INSERTION LIMIT. THE MOST LIKELY ROOT CAUSE WAS ATTRIBUTED TO A DESIGN ISSUE THAT ALLOWS THE SCREWS TO BE PULLED THROUGH THE PLATE HOLES WHEN TIGHTENED AT > 30-DEGREES OFF-AXIS. AS CORRECTIVE ACTION, THE NON-LOCKING SCREW WAS REDESIGNED TO A LOCKING SCREW AND THE PREVIOUS DESIGN WAS DISCONTINUED. PER ENGINEERING CHANGE ORDER (EO-00012), NO FURTHER ACTION WAS REQUIRED. THIS DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THERE IS NO ADDITIONAL INFORMATION ON THE EVENT, PATIENT, OR DEVICES. CONCOMITANT DEVICES: 0MM PLATE -2000-2000 SN: (B)(4). 2001-3518 18MM 3.5 NONLOCKING SCREW SN; (B)(4). 2.7 DRILL BIT 2100-2700 SN: (B)(4). 2001-3518 18MM 3.5 NONLOCKING SCREW SN; (B)(4). 2.7 DRILL BIT 2100-2700 SN: (B)(4).

Description of Event or Problem · 0

AS PART OF DUE DILIGENCE CONDUCTED DURING A COMPANY ACQUISITION AUDIT THIS COMPLAINT WAS DETERMINED TO BE REPORTABLE FOR FDA MDR. THIS MEDWATCH IS BEING REPORTED BY THIS COMPANY ON BEHALF OF THE LISTED MANUFACTURER FOUND IN D3. IT WAS REPORTED THAT A PATIENT EXPERIENCED A SURGERY FOR A LATERAL COLUMN LENGTHENING OF THE CALCANEUS ON (B)(6) 2016. DURING THE PROCEDURE THE PLATE WAS PLACED OVER A CADAVER GRAFT IN THE CALCANEUS TO FIXATE THE WEDGE IN THE PLATE. THE PLATE WAS CHOSEN BY THE SURGEON AND PLACED IN THE PATIENT. ONE 3.5 NON-LOCKING SCREW WAS PLACED IN THE ANTERIOR PROCESS OF THE CALCANEUS THROUGH THE PLATE AFTER BEING DRILLED WITH A 2.7MM DRILL. THE NEXT SCREW WAS DRIVEN ON THE OPPOSITE SIDE TO THE OSTEOTOMY WITH THE 2.7MM DRILL AND THEN A 3.5 MM NON-LOCKING SCREW WAS SCREWED INTO THE PLATE. THE SCREW WAS RATCHETED DOWN WITH A HAND DRIVER AND CONTINUED THROUGH THE PLATE COMPLETELY. THE SURGEON REMOVED THE PLATE AND SCREW TO REPOSITION THE PLATE. THE SURGEON PUT ANOTHER SCREW IN THE PATIENT WHICH WENT THROUGH THE PLATE AGAIN. THE SCREW WENT THROUGH THE INFERIOR PLATE HOLE THIS TIME, THIS OCCURRED IN TWO DIFFERENT SCREW HOLES. ALTERNATIVE HARDWARE FROM ANOTHER MANUFACTURER WAS USED. THE PATIENT ENDURED 45 MINUTES OF ADDITIONAL ANESTHESIA TIME DUE TO THE DELAY IN SURGERY. THERE IS NO ADDITIONAL INFORMATION ON THE EVENT, PATIENT, OR DEVICES. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER REPORT NUMBERS ARE 1038671-2019-00001 AND 1038671-2019-00002.

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICES WERE NOT RETURNED FOR ANALYSIS. A PULL THROUGH TEST WAS CONDUCTED USING SIMILAR PARTS. THE SCREWS WERE TIGHTENED AGAINST THE PLATE ON-AXIS (+/- 12.5-DEGREES), AT 20-DEGREES OFF-AXIS, AND 30-DEGREES OFF-AXIS. FURTHER TESTING WAS CONDUCTED USING THE NEWLY DESIGNED LOCKING SCREWS. THE MOST LIKELY ROOT CAUSE WAS ATTRIBUTED TO A DESIGN ISSUE THAT ALLOWS THE SCREWS TO BE PULLED THROUGH THE PLATE HOLES WHEN TIGHTENED AT > 30-DEGREES OFF-AXIS. AS CORRECTIVE ACTION, THE NON-LOCKING SCREW WAS REDESIGNED TO A LOCKING SCREW AND THE PREVIOUS DESIGN WAS DISCONTINUED. PER ENGINEERING CHANGE ORDER ((B)(4)), NO FURTHER ACTION WAS REQUIRED. THIS DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THERE IS NO ADDITIONAL INFORMATION ON THE EVENT, PATIENT, OR DEVICES. CONCOMITANT MEDICAL PRODUCTS: 0MM PLATE -2000-2000 SN: (B)(4). 2001-3518 18MM 3.5 NONLOCKING SCREW SN: (B)(4). 2.7 DRILL BIT 2100-2700 SN: (B)(4).

Description of Event or Problem · 1

AS PART OF DUE DILIGENCE CONDUCTED DURING A COMPANY ACQUISITION AUDIT THIS COMPLAINT WAS DETERMINED TO BE REPORTABLE FOR FDA MDR. THIS MEDWATCH IS BEING REPORTED BY THIS COMPANY ON BEHALF OF THE LISTED MANUFACTURER. IT WAS REPORTED THAT A PATIENT EXPERIENCED A SURGERY FOR A LATERAL COLUMN LENGTHENING OF THE CALCANEUS ON (B)(6) 2016. DURING THE PROCEDURE THE PLATE WAS PLACED OVER A CADAVER GRAFT IN THE CALCANEUS TO FIXATE THE WEDGE IN THE PLATE. THE PLATE WAS CHOSEN BY THE SURGEON AND PLACED IN THE PATIENT. ONE 3.5 NON-LOCKING SCREW WAS PLACED IN THE ANTERIOR PROCESS OF THE CALCANEUS THROUGH THE PLATE AFTER BEING DRILLED WITH A 2.7MM DRILL. THE NEXT SCREW WAS DRIVEN ON THE OPPOSITE SIDE TO THE OSTEOTOMY WITH THE 2.7MM DRILL AND THEN A 3.5 MM NON-LOCKING SCREW WAS SCREWED INTO THE PLATE. THE SCREW WAS RATCHETED DOWN WITH A HAND DRIVER AND CONTINUED THROUGH THE PLATE COMPLETELY. THE SURGEON REMOVED THE PLATE AND SCREW TO REPOSITION THE PLATE. THE SURGEON PUT ANOTHER SCREW IN THE PATIENT WHICH WENT THROUGH THE PLATE AGAIN. THE SCREW WENT THROUGH THE INFERIOR PLATE HOLE THIS TIME, THIS OCCURRED IN TWO DIFFERENT SCREW HOLES. ALTERNATIVE HARDWARE FROM ANOTHER MANUFACTURER WAS USED. THE PATIENT ENDURED 45 MINUTES OF ADDITIONAL ANESTHESIA TIME DUE TO THE DELAY IN SURGERY. THERE IS NO ADDITIONAL INFORMATION ON THE EVENT, PATIENT, OR DEVICES. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER REPORT NUMBERS ARE 1038671-2019-00001 AND 1038671-2019-00002.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
539921 EPIC PLATE SYSTEM BONE FIXATION PLATE, SCREW HRS EPIC EXTREMITY

Patients

Seq Age Sex Outcome Treatment
1 42 YR Required Intervention 2.7 DRILL BIT 2100-2700 SN: N658342| 2001-3518 18MM 3.5 NONLOCKING SCREW SN, CWO2O6| SEE H10.| SEE H10.| 2.7 DRILL BIT 2100-2700 SN: (B)(4)| 2001-3518 18MM 3.5 NONLOCKING SCREW SN: (B)(4)