ABL800 FLEX
Report
- Report Number
- 3002807968-2019-00028
- Event Type
- Malfunction
- Date Received
- June 28, 2019
- Date of Event
- June 13, 2019
- Report Date
- May 19, 2020
- Manufacturer
- RADIOMETER MEDICAL APS
- Product Code
- JFP
- UDI-DI
- 05700693938011
- PMA / PMN Number
- K041874
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THE MEASUREMENT ERROR WAS CAUSED BY A FAILING CREA MEMBRANE. THE CUSTOMER HAD NOT FOLLOWED THE RECOMMENDED PROCEDURE FOR QUALITY CONTROLS FOR MEASUREMENTS OF CREA. FURTHERMORE, THERE WAS FOUND A UNDERLYING INSTABILITY IF THE CREA ELECTRODE, WHICH WAS NOT DETECTED AS THE CUSTOMER HAD DISABLED THE ALARMS.
A FIELD SERVICE ENGINEER REPLACED THE AFFECTED ANALYZER AT THE HOSPITAL WITH AN ALTERNATIVE ANALYZER.
ON (B)(6) 2019, A CREA MEASUREMENT MEASURED ON AN ABL801 WAS WITHIN THE NORMAL RANGE AND THE PATIENT WAS PROVIDED CONTRAST AGENT FOR AN IMAGE DIAGNOSIS. HOWEVER, THE PATIENT ORIGINALLY HAD A POOR KIDNEY FUNCTION. THE DOCTOR CONFIRMED THAT THE TAKEN CONTRAST AGENT DID NOT DAMAGE THE PATIENT. THE PATIENT HAS NOT RECEIVED ANY DAMAGE AND ANY TREATMENT BASED ON THE CONTRAST AGENT. A MEASUREMENT OF 0.9 MG/DL WAS MEASURED ON THE ABL801 AND A COMPARISON MEASUREMENT IN THE LABORATORY GAVE 3.5 MG/DL. HOWEVER, ALL LEVELS OF QUALITY CONTROLS (QCS) HAD BEEN USED, DESPITE THAT CUSTOMERS HAS BEEN INSTRUCTED TO NARROW THE RANGE AND USE QC LEVEL 1. FURTHERMORE, THE ALERT FOR QC OUT OF RANGE WAS NOT SET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 536569 | ABL800 FLEX | ABL800 FLEX | JFP | RADIOMETER MEDICAL APS | 393-801 | 05700693938011 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |