FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722CAS

MDR report key: 8743302 · Received June 28, 2019

Report

Report Number
2032227-2019-22645
Event Type
Malfunction
Date Received
June 28, 2019
Date of Event
June 23, 2019
Report Date
October 10, 2019
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169919389
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE POWER UP PROPERLY AFTER BATTERY INSTALLATION. NO BLANK DISPLAY NOTED. HOWEVER, DEVICE RECEIVED WITH FROZEN SCREEN STUCK AT NUMBER 2, PROBLEM ISOLATED TO MOTHER BOARD. UNABLE TO PERFORM THE SELF-TEST, A21 ERROR TEST, DISPLACEMENT TEST DUE TO FROZEN SCREEN.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A BLANK DISPLAY. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND THE DISPLAY DID NOT RETURN. CUSTOMER STATES BATTERY COMPARTMENT IS NEITHER DAMAGED NOR CORRODED. CUSTOMER STATES THE SPRING IS NEITHER DAMAGED NOR CORRODED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE CUSTOMER WAS ADVISED THE PUMP WILL BE REPLACED AS PER TROUBLESHOOT. THE INSULIN PUMP WILL BE RETURN FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
536794 PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722CAS PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722CAS B4722CASJ 00643169919389

Patients

Seq Age Sex Outcome Treatment
1