FDA Adverse Event
Malfunction
Summary report: N
BRILLIANCE EBW WORKSTATION CARDIAC VIEWER
MDR report key: 874065
·
Received June 29, 2007
Report
- Report Number
- 1525965-2007-00005
- Event Type
- Malfunction
- Date Received
- June 29, 2007
- Date of Event
- May 30, 2007
- Report Date
- June 29, 2007
- Manufacturer
- PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
- Product Code
- JAK
- PMA / PMN Number
- K982060
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE EXTENDED BRILLIANCE WORKSPACE (EBW) HAS AN OPTIONAL SOFTWARE APPLICATION FUNCTION CALLED THE CARDIAC VIEWER. IT IS POSSIBLE FOR AN SET OF IMAGES TO BE EXPANDED OR ZOOMED AND SAVE AS A NEW SET OF IMAGES. WHEN THE SAVED SET OF IMAGES ARE VIEWED USING A FUNCTION CALLED THE CT VIEWER, OR ARE TRANSFERRED TO A PACS (PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM), MEASUREMENTS MADE ON THESE ZOOMED IMAGES ARE INCORRECT. THEY MEASUREMENT ERROR IS PROPORTIONAL TO THE ZOOM FACTOR. AN IMAGE WHICH IS ZOOMED BY A FACTOR OF TWO WILL SHOW DISTANCE MEASUREMENTS THAT ARE INCORRECT AND ENLARGED BY A FACTOR OR TWO.