FDA Adverse Event Malfunction Summary report: N

BRILLIANCE EBW WORKSTATION CARDIAC VIEWER

MDR report key: 874065 · Received June 29, 2007

Report

Report Number
1525965-2007-00005
Event Type
Malfunction
Date Received
June 29, 2007
Date of Event
May 30, 2007
Report Date
June 29, 2007
Manufacturer
PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
Product Code
JAK
PMA / PMN Number
K982060
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE EXTENDED BRILLIANCE WORKSPACE (EBW) HAS AN OPTIONAL SOFTWARE APPLICATION FUNCTION CALLED THE CARDIAC VIEWER. IT IS POSSIBLE FOR AN SET OF IMAGES TO BE EXPANDED OR ZOOMED AND SAVE AS A NEW SET OF IMAGES. WHEN THE SAVED SET OF IMAGES ARE VIEWED USING A FUNCTION CALLED THE CT VIEWER, OR ARE TRANSFERRED TO A PACS (PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM), MEASUREMENTS MADE ON THESE ZOOMED IMAGES ARE INCORRECT. THEY MEASUREMENT ERROR IS PROPORTIONAL TO THE ZOOM FACTOR. AN IMAGE WHICH IS ZOOMED BY A FACTOR OF TWO WILL SHOW DISTANCE MEASUREMENTS THAT ARE INCORRECT AND ENLARGED BY A FACTOR OR TWO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BRILLIANCE EBW WORKSTATION CARDIAC VIEWER VIEWING WORKSTATION FOR CT IMAGES JAK PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.