FDA Adverse Event
Malfunction
Summary report: N
FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER
MDR report key: 8739528
·
Received June 26, 2019
Report
- Report Number
- MW5087643
- Event Type
- Malfunction
- Date Received
- June 26, 2019
- Date of Event
- June 13, 2019
- Report Date
- June 24, 2019
- Manufacturer
- HUDSON RCI, TELEFLEX MEDICAL
- Product Code
- BTM
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- PA, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
FLOW INFLATING NEONATE AMBU BAG WOULD NOT MAINTAIN PRESSURE DURING POSITIVE PRESSURE VENTILATION. NEW AMBU BAG DEFECTIVE X3 BEFORE FUNCTIONING BAG OBTAINED. NOTED THAT THE WHITE PLASTIC RING AT THE TOP OF THE GREEN FLOW INFLATING BAG WHERE IT MEETS THE CONNECTION THE MANOMETER WAS CRACKED. ENTIRE INVENTORY WAS CHECKED AND AMONG DOZENS OF PRODUCT 15 FROM 4 DIFFERENT LOT NUMBERS WERE FOUND TO BE CRACKED. DISTRIBUTOR WAS CONTACTED AND DEFECTIVE ITEMS WERE REMOVED FROM STOCK. LOT NUMBERS AND NUMBER OF THAT LOT NUMBER THAT WERE FOUND DEFECTIVE: 316763 - 1 BAG, 317232 - 3 BAGS, 317645 - 6 BAGS, 318787 - 5 BAGS. FDA SAFETY REPORT ID# (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 531636 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 318787 | ||
| 531640 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 316763 | ||
| 531641 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317645 | ||
| 531642 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317645 | ||
| 531643 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317232 | ||
| 531644 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317232 | ||
| 531645 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317232 | ||
| 531646 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317645 | ||
| 531647 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317645 | ||
| 531648 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317645 | ||
| 531649 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 317645 | ||
| 531650 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 318787 | ||
| 531651 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 318787 | ||
| 531652 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 318787 | ||
| 531653 | FLOW INFLATING INFANT AMBU BAG 1/2L HYPERINFLATION SYSTEM WITH MANOMETER | VENTILATOR, EMERGENCY, MANUAL (RESUSCITATION) | BTM | HUDSON RCI, TELEFLEX MEDICAL | 318787 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 DA |