FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 8737359 · Received June 26, 2019

Report

Report Number
1645337-2019-14486
Event Type
Injury
Date Received
June 26, 2019
Date of Event
April 1, 2019
Report Date
June 7, 2019
Manufacturer
MENTOR TEXAS
Product Code
FTR
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 7/30/2019, MENTOR WAS MADE AWARE THE PATIENT WOULD UNDERGO EXPLANTATION ON (B)(6) 2019. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 9/23/2019, MENTOR RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE PATIENT'S DEVICES WERE CONFIRMED TO BE GEL MENTOR MEMORYGEL BREAST IMPLANT 325CC, CATALOG# 3507325BC, LOT# 5924379 (LEFT), LOT# 5919211 (RIGHT). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON (B)(6) 2019, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THE PATIENT HAD ALSO EXPERIENCED A RUPTURE ON THE RIGHT SIDE DEVICE POST-OPERATIVELY. MENTOR WAS ALSO MADE AWARE THE PATIENT'S PREVIOUSLY UNKNOWN DEVICES WERE SALINE MENTOR SMOOTH ROUND MODERATE PLUS PROFILE 350CC BREAST IMPLANTS, CATALOG# 3502350, LOT# 5982470, SERIAL# (B)(4). A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 11/06/2019, MENTOR COMPLETED EVALUATION OF THE DEVICE. DEVICE EVALUATION SUMMARY: DURING EVALUATION OF THE SAMPLE WAS FOUND RUPTURED. CREASE WAS FOUND IN THE EDGES OF THE TEAR. NO OTHER ANOMALIES WERE DISCOVERED. THESE KIND OF FINDINGS ARE CONSISTENT WITH A CREASE/FOLD FAILURE WHICH IS A KNOWN INHERENT RISK ASSOCIATED WITH THE USE OF GEL-FILLED MAMMARY PROSTHESES AND MAY BE THE RESULT OF ONE OR MORE OF THE FOLLOWING CONTRIBUTING FACTORS: CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE - SUCH AS TOO SMALL A BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. BASED ON THE INFORMATION REPORTED AND/OR THE PRODUCT INVESTIGATION, THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH THE MANUFACTURING PROCESS. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION REVIEW COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE HAS BEEN RECEIVED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6)-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH UNSPECIFIED MENTOR GEL IMPLANTS AND EXPERIENCED A RUPTURE ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED BY A PHYSICIAN. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
529905 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 5924379

Patients

Seq Age Sex Outcome Treatment
1 37 YR Other| R