MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2019-14486
- Event Type
- Injury
- Date Received
- June 26, 2019
- Date of Event
- April 1, 2019
- Report Date
- June 7, 2019
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
ON 7/30/2019, MENTOR WAS MADE AWARE THE PATIENT WOULD UNDERGO EXPLANTATION ON (B)(6) 2019. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON 9/23/2019, MENTOR RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE PATIENT'S DEVICES WERE CONFIRMED TO BE GEL MENTOR MEMORYGEL BREAST IMPLANT 325CC, CATALOG# 3507325BC, LOT# 5924379 (LEFT), LOT# 5919211 (RIGHT). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON (B)(6) 2019, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THE PATIENT HAD ALSO EXPERIENCED A RUPTURE ON THE RIGHT SIDE DEVICE POST-OPERATIVELY. MENTOR WAS ALSO MADE AWARE THE PATIENT'S PREVIOUSLY UNKNOWN DEVICES WERE SALINE MENTOR SMOOTH ROUND MODERATE PLUS PROFILE 350CC BREAST IMPLANTS, CATALOG# 3502350, LOT# 5982470, SERIAL# (B)(4). A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON 11/06/2019, MENTOR COMPLETED EVALUATION OF THE DEVICE. DEVICE EVALUATION SUMMARY: DURING EVALUATION OF THE SAMPLE WAS FOUND RUPTURED. CREASE WAS FOUND IN THE EDGES OF THE TEAR. NO OTHER ANOMALIES WERE DISCOVERED. THESE KIND OF FINDINGS ARE CONSISTENT WITH A CREASE/FOLD FAILURE WHICH IS A KNOWN INHERENT RISK ASSOCIATED WITH THE USE OF GEL-FILLED MAMMARY PROSTHESES AND MAY BE THE RESULT OF ONE OR MORE OF THE FOLLOWING CONTRIBUTING FACTORS: CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE - SUCH AS TOO SMALL A BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. BASED ON THE INFORMATION REPORTED AND/OR THE PRODUCT INVESTIGATION, THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH THE MANUFACTURING PROCESS. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION REVIEW COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE HAS BEEN RECEIVED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6)-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH UNSPECIFIED MENTOR GEL IMPLANTS AND EXPERIENCED A RUPTURE ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED BY A PHYSICIAN. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 529905 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 5924379 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | Other| R |