FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,10

MDR report key: 8735888 · Received June 26, 2019

Report

Report Number
0002023141-2019-00336
Event Type
Injury
Date Received
June 26, 2019
Date of Event
May 29, 2019
Report Date
July 11, 2019
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019225
PMA / PMN Number
K072589
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

SIMILAR COMPLAINTS FOR IMPLANT INFECTION HAVE BEEN PREVIOUSLY INVESTIGATED. REFER TO ATTACHED SUMMARY INVESTIGATION. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORD REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH DID NOT SUGGEST THE LIKELY RELEASE OF NONCONFORMING PRODUCT. ZIMMER BIOMET QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1223103). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1223103) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: DATE OF THIS REPORT. CHECKED "FOLLOW-UP." CHECKED FOLLOW-UP TYPE. DEVICE EVALUATED BY MANUFACTURER. ENTERED EVALUATION CODES. ADDED MANUFACTURER NARRATIVE.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING SUBMITTED TO REPORT (B)(4). THE FOLLOWING INFORMATION IS UNKNOWN AT THE TIME OF THIS REPORT: DATE OF BIRTH NOT PROVIDED. WEIGHT, ETHNICITY NOT PROVIDED. THE REPORTED DEVICE HAS BEEN RECEIVED FOR EVALUATION. INVESTIGATION HAS NOT YET BEGUN. ONCE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED WITH INFECTION AT TOOTH LOCATION 8. THE IMPLANT WAS REMOVED. THERE WAS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
527918 IMP,TSV,4.1MM,SBM,10 DENTAL IMPLANT DZE ZIMMER DENTAL N/A 1223103 00889024019225

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention