AMPLICOR CT/NG TEST
Report
- Report Number
- 2243471-2006-00011
- Event Type
- Malfunction
- Date Received
- September 28, 2006
- Report Date
- September 13, 2006
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC.
- Product Code
- MKZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE PACKAGE INSERT THAT ACCOMPANIES THE AMPLICOR CT/NG TEST STATES IN THE INSTRUCTIONS FOR USE SECTION, PART D AMPLIFICATION, STEP 3 "DURING RUN SET-UP, SET THE REACTION VOLUME TO 100UL BY FIRST CHANGING THE RAMP SPEED FROM THE 9600 MODE TO THE MAX MODE. TO DO THIS, CURSOR DOWN TO RAMP SPEED AND SELECT MAX". APPLIED BIOSYSTEMS GENEAMP PCR SYSTEM 9700 THERMAL CYCLERS WITH THE ALUMINUM HEAT BLOCK WILL NOT ACCEPT THIS SETTING. ONLY THERMAL CYCLERS EQUIPPED WITH THE GOLD-PLATED HEAT BLOCK WILL ACCEPT THE RAMP SPEED SETTING OF MAX. THE PURPOSE OF PROVIDING THE PRODUCT BULLETIN WAS TO PROVIDE ADDITIONAL INFORMATION ON THIS SETTING REQUIREMENT. ADD'L LOT # G14950.
IN JUNE 2006, ROCHE MOLECULAR SYSTEMS, INC. PROVIDED A PRODUCT BULLETIN CLARIFYING THE USE OF THE GOLD PLATED HEAT BLOCK AND THE ALUMINUM HEAT BLOCK ON THE APPLIED BIOSYSTEMS GENEAMP PCR SYSTEM 9700 THERMAL CYCLER. IN RESPONSE TO THE THE BULLETIN, THE CUSTOMER REPORTED THEY HAD BEEN USING THE ALUMINUM HEAT BLOCK WHEN PERFORMING THE AMPLICOR CT/NG TEST. THE ALUMINUM HEAT BLOCK WILL NOT ACCEPT A MAXIMUM RAMP SPEED WHICH IS REQUIRED TO PERFORM THE AMPLICOR CT/NG TEST. THE CURRENT AMPLICOR CT/NG TEST PACKAGE INSERT DESCRIBES THE SETTING OF THE TEMPERATURE RAMP SPEED TO MAXIMUM BUT DOES NOT IDENTIFY THE HEAT BLOCK TO BE USED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMPLICOR CT/NG TEST | IN VITRO DIAGNOSTIC NUCLEIC ACID TEST | MKZ | ROCHE MOLECULAR SYSTEMS, INC. | NA | H02039 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |