GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2019-60722
- Event Type
- Malfunction
- Date Received
- June 25, 2019
- Date of Event
- May 28, 2019
- Report Date
- July 26, 2019
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC
- Product Code
- GEX
- UDI-DI
- 00878953005515
- PMA / PMN Number
- K120870
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICES MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. ANALYSIS OF THE DEVICE REVEALED THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION AT OUTPUT WINDOW. THE METAL CAP EXHIBITS SEVERE CHARRED DETRITUS ADHESION ON SURFACE. CAP WEAR IS A KNOWN INHERENT RISK OF DEVICE. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER. THE IDENTIFIED ISSUES NOTED ABOVE MAY ACTIVATE THE FIBERLIFE FUNCTION WHICH WOULD MODULATE THE POWER SHOWING A PULSING BEAM OR SYSTEM WOULD BE PLACED INTO STANDBY MODE. THIS WOULD CAUSE REDUCED VAPORIZATION EFFICIENCY. IN ADDITION, ANALYSIS IDENTIFIED THE GLASS CAP HAS A CIRCUMFERENTIAL FRACTURE ON THE DISTAL SIDE OF FIBER/CAP FUSION ZONE AT THE BEVEL EDGE. THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF OUTER FLOW TUBING. THE OUTER FLOW TUBING OPEN END EXHIBITS MINOR SCRATCH MARKS. THE POTENTIAL FOR FORWARD FIRING MAY EXIST. DUE TO THE OBSERVED GLASS CAP DISTAL FRACTURE, AN EVALUATION CONCLUSION CODE OF DESIGN INADEQUATE FOR PURPOSE OF THE DEVICE WAS ASSIGNED TO THIS INVESTIGATION. IT IS PROBABLE THAT A TEMPERATURE HIGHER OR CLOSE TO EPOXY DEGRADATION TEMPERATURE NEAR THE LASER BEAM OUTPUT WINDOW MAY BE A MAJOR IMPACTING FACTOR LEADING TO EPOXY FAILURE AND SUBSEQUENT FIBER BREAKAGE. TISSUE ADHESION FROM CONSTANT AND HEAVY TISSUE CONTACT COULD BE THE CAUSE RESULTING IN THE OBSERVED CIRCUMFERENTIAL FRACTURE.
IT WAS REPORTED THAT THE LASER BEAM WAS COMING STRAIGHT OUT OF THE TIP OF FIBER AFTER 180497 JOULES OF USE. THE FIBER TIP WAS NOT CLEANED DURING THE PROCEDURE. THE CASE WAS COMPLETE WITH ANOTHER FIBER AND THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.
IT WAS REPORTED THAT THE LASER BEAM WAS COMING STRAIGHT OUT OF THE TIP OF FIBER AFTER 180497 JOULES OF USE. THE FIBER TIP WAS NOT CLEANED DURING THE PROCEDURE. THE CASE WAS COMPLETE WITH ANOTHER FIBER AND THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 525907 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC | 0010-2400 | 0023140183 | 00878953005515 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |