FDA Adverse Event Malfunction Summary report: N

HEMOCUE HB 201+ SYSTEM

MDR report key: 8730106 · Received June 25, 2019

Report

Report Number
3003044483-2019-00009
Event Type
Malfunction
Date Received
June 25, 2019
Date of Event
May 16, 2019
Report Date
June 25, 2019
Manufacturer
HEMOCUE AB
Product Code
GKR
PMA / PMN Number
K032203
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ANALYZER WAS SENT TO HEMOCUE AB WHERE THE CUSTOMER PROBLEM WAS CONFIRMED, IT WAS DISCOVERED THAT THE EEPROM MEMORY HAD BEEN CORRUPTED. THE ROOT CAUSE WAS FOUND TO BE A GLITCH IN THE CONNECTION BETWEEN COMPONENTS ON THE MAIN BOARD WHICH HAS CAUSED READ AND WRITE ERRORS LEADING TO CORRUPT VALUES IN EEPROM MEMORY. THE ERROR IN EEPROM MEMORY WILL BE PERMANENT IF THE ROOT CAUSE IS HAPPENING IN COMBINATION WITH OTHER IDENTIFIED SCENARIOS LEADING TO POTENTIAL INCORRECT MEASUREMENTS VALUES. CORRECTIVE ACTION IS TO IMPROVE THE SOLDERING PROCESS OF COMPONENTS ON THE MAIN BOARD. NOTE, WHEN THE ROOT CAUSE IS NOT OCCURRING IN A COMBINATION WITH THE IDENTIFIED SCENARIOS IT WILL LEAD TO AN ERROR CODE AND NO RESULTS WILL BE DISPLAYED.

Description of Event or Problem · 1

HEMOCUE AB RECEIVED A COMPLAINT THAT THE ANALYZER HAD A FAULTY SWITCH AND DID NOT START.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
523638 HEMOCUE HB 201+ SYSTEM HEMOGLOBIN TEST SYSTEM GKR HEMOCUE AB 121704

Patients

Seq Age Sex Outcome Treatment
1