FDA Adverse Event Injury Summary report: N

HERO

MDR report key: 8729509 · Received June 24, 2019

Report

Report Number
3014660737-2019-90001
Event Type
Injury
Date Received
June 24, 2019
Date of Event
May 26, 2019
Report Date
June 24, 2019
Manufacturer
HERO HEALTH, INC.
Product Code
NXB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

USER DISPENSED OVER 12 UNSCHEDULED PILLS BY PRESSING THE DISPENSE BUTTON REPEATEDLY. THE FOLLOWING DAY, A CAREGIVER TOOK THE USER TO THE ER. PER USER'S CAREGIVER, USER IS EXPECTED TO MAKE A FULL RECOVERY. PER THE DEVICE INSTRUCTION MANUAL, THE DEVICE IS NOT INTENDED TO REPLACE AN ATTENTIVE CAREGIVER, SHOULD ONE BE REQUIRED BY THE USER. THE USER'S CAREGIVER DID NOT SET UP A CAREGIVER ACCOUNT, WHICH WOULD HAVE ALERTED THE CAREGIVER WITH AN APP NOTIFICATION WHEN THE FIRST UNSCHEDULED PILL WAS DISPENSED BY THE END USER. THE EVENT WAS CAUSED BY USER ERROR, IN THAT THE REDUNDANT MEANS TO PREVENT OR MITIGATE THIS TYPE OF EVENT - DISPENSING MULTIPLE UNSCHEDULED MEDICATIONS - WERE PURPOSELY NOT UTILIZED BY THE PATIENT'S CAREGIVER WHO SET UP THE DEVICE.

Description of Event or Problem · 1

USER DISPENSED OVER 12 UNSCHEDULED PILLS BY PRESSING THE DISPENSE BUTTON REPEATEDLY. THE FOLLOWING DAY, A CAREGIVER TOOK THE USER TO THE ER. PER USER'S CAREGIVER, USER IS EXPECTED TO MAKE A FULL RECOVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
522316 HERO HERO DISPENSER - SOLID MEDICATION NXB HERO HEALTH, INC. H-100

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization