FDA Adverse Event Malfunction Summary report: N

MONOSYN VIOLET 6/0 (0.7) 70CM HR13 (M)

MDR report key: 8727702 · Received June 24, 2019

Report

Report Number
3003639970-2019-00489
Event Type
Malfunction
Date Received
June 24, 2019
Date of Event
April 16, 2019
Report Date
June 24, 2019
Manufacturer
B.BRAUN SURGICAL SA
Product Code
GAM
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN SURGICAL S.A. (MANUFACTURER). EXEMPTION NUMBER: E2014012. REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K011375. INVESTIGATION: SAMPLES RECEIVED: 2 OPEN UNITS. ANALYSIS AND RESULTS: AS NO BATCH NUMBER IS AVAILABLE, THE BATCH MANUFACTURING RECORD CANNOT BE REVIEWED. WE HAVE RECEIVED TWO OPEN AND UNUSED SAMPLES WITHOUT ALUMINIUM PACK. THE NEEDLE IS DETACHED FROM THE THREAD IN BOTH SAMPLES RECEIVED AND THE THREAD IS STILL WOUND IN THE CARDBOARD SUPPORT. FINAL CONCLUSION: TAKING INTO ACCOUNT THAT THE SAMPLES RECEIVED DO NOT FULFIL B. BRAUN SURGICAL SPECIFICATIONS, WE CONCLUDE THAT THE COMPLAINT IS CONFIRMED BY EVIDENCE OF THE SAMPLES RECEIVED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. NO CORRECTIVE/PREVENTIVE ACTIONS NEEDED.

Description of Event or Problem · 0

IT WAS REPORTED THE NEEDLE AND THREAD SEPARATED. THE REPORTER INDICATED THAT WHEN THE PACKAGE WAS OPENED, IT WAS FOUND THAT THE NEEDLE AND THREAD HAD DETACHED. THIS WAS FOUND PRIOR TO USE. ADDITIONAL INFORMATION HAS NOT BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
520430 MONOSYN VIOLET 6/0 (0.7) 70CM HR13 (M) SYNTHETIC ABSORBABLE MONOFILAR GAM B.BRAUN SURGICAL SA C0022003

Patients

Seq Age Sex Outcome Treatment
1