630G INSULIN PUMP MMT-1715K 630G BLACK MG
Report
- Report Number
- 2032227-2019-20416
- Event Type
- Malfunction
- Date Received
- June 24, 2019
- Date of Event
- June 15, 2019
- Report Date
- August 19, 2019
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- UDI-DI
- 00763000190439
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
DEVICE PASSED SELFTEST, REWIND, SEATING, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST AND DISPLACEMENT TEST. NO UNEXPECTED INSULIN FLOW BLOCKED ALARMS NOTED. DEVICE RECEIVED WITH THE INSULIN FLOW BLOCKED ALARM FUNCTIONING PROPERLY. (B)(4).
THE CUSTOMER REPORTED VIA PHONE CALL THAT INSULIN PUMP HAD FLOW BLOCK ALARM. BLOOD GLUCOSE WAS UNKNOWN. TROUBLESHOOTING WAS PERFORMED. CUSTOMER STATES THEY HAVE TRIED ALL REMEDIES BUT SAME ISSUE WAS REOCCURRED. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 519955 | 630G INSULIN PUMP MMT-1715K 630G BLACK MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1715K | HG2W2YN | 00763000190439 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |