CONTOUR NEXT
Report
- Report Number
- 1810909-2019-00260
- Event Type
- Malfunction
- Date Received
- June 21, 2019
- Date of Event
- May 18, 2019
- Report Date
- May 23, 2019
- Manufacturer
- ASCENSIA DIABETES CARE US INC.
- Product Code
- NBW
- UDI-DI
- 00301939697010
- PMA / PMN Number
- K121190
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER DID NOT RETURN THE SUSPECTED PRODUCT. AN IN-HOUSE TESTING WAS PERFORMED USING THE IN-HOUSE CONTOUR NEXT METER WITH IN-HOUSE CONTOUR NEXT TEST STRIPS FROM LOT # 8FPEG30A, WHICH GAVE SATISFACTORY PERFORMANCE.
THE PATIENT/FAMILY WAS THE INITIAL REPORTER, SO PERSONAL INFORMATION WAS NOT ENTERED. THIS EVENT IS RELATED TO MDRS 1810909-2019-00261, 1810909-2019-00262, 1810909-2019-00263, AND 1810909-2019-00264.
BASED ON THE INFORMATION RECEIVED, CUSTOMER USED CONTOUR NEXT TEST STRIPS LOT # 8FPEG30A FOR THIS EVENT. DATE RECEIVED BY MANUFACTURER HAVE BEEN CORRECTED TO REFLECT THE CORRECT PRODUCT INFORMATION.
AT 11:09 A.M., THE CUSTOMER OBTAINED A BLOOD GLUCOSE READING OF 305 MG/DL ON A CONTOUR NEXT METER. THE CUSTOMER HAD NO SYMPTOMS OF HYPERGLYCEMIA. WITHIN 10 MINUTES, THE CUSTOMER PERFORMED A REPEAT TESTING ON THE BREEZE2 METER AND OBTAINED A READING OF 40-50% LOWER COMPARED TO THAT OBTAINED ON THE CONTOUR NEXT METER. THE SPECIFIC READING OBTAINED ON THE BREEZE2 METER WAS NOT PROVIDED. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE TEST STRIPS FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 514663 | CONTOUR NEXT | BLOOD GLUCOSE TEST STRIPS | NBW | ASCENSIA DIABETES CARE US INC. | 7312 | 8FPEG30A | 00301939697010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |