BACT/ALERT MP CULTURE BOTTLES
Report
- Report Number
- 3002769706-2007-00004
- Event Type
- Other
- Date Received
- June 26, 2007
- Date of Event
- June 1, 2007
- Report Date
- June 22, 2007
- Manufacturer
- BIOMERIEUX, INC.
- Product Code
- MDB
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- UNKNOWN
Narratives
THE BACT/ALERT MP PROCESS SYSTEM CONSIST OF THE BACT/ALERT MP PROCESS BOTTLE WITH A REMOVABLE CLOSURE USED IN CONJUNCTION WITH THE MB/BACT/ALERT ANTIBIOTIC SUPPLEMENT (AND /OR THE MB/BACT/ALERT 3D MYCOBACTERIA DETECTION SYSTEMS FOR THE RECOVERY AND DETECTION OF MYCOBACTERIA FROM STERILE BODY SPECIMENS OTHER THAN BLOOD) AND FROM DIGESTED-DECONTAMINATED CLINICAL SPECIMENS. THERE IS A POTENTIAL FOR AN ADVERSE EVENT WERE THIS TO RECUR AND THE INTEGRITY OF A POSITIVE BOTTLE WERE TO BE COMPROMISED. IN THE EVENT OF BREAKAGE, THE USER COULD BE EXPOSED TO AEROSOLIZED MYCOBACTERIA SAMPLES THAT CAN HAVE SIGNIFICANT HEALTH EFFECTS. CURRENTLY, THIS ISSUE IS SPECIFIC TO THIS CUSTOMER ONLY THE CUSTOMER INDICATED THAT THEY RECENTLY CHANGED THE DISINFECTANT USED IN THERE PROCESSING OF BOTTLES. THEY HAD NOT NOTICED THIS ISSUE OF CRACKED BOTTLES PRIOR TO THE CHANGE IN DISINFECTANT. THEREFORE, BIOMERIEUX IS INVESTIGATING THE INTERACTION OF THIS DISINFECTANT WITH THE MP BOTTLE AS PART OF THE OVERALL INVESTIGATION INTO THIS COMPLAINT.
INOCULATED BACT/ALERT MP PLASTIC BOTTLES REMOVED FROM THE INCUBATOR HAD "CRACKS" AROUND THE NECK OF THE BOTTLES. IT DOES NOT APPEAR THAT THE INTEGRITY OF THE BOTTLES HAD BEEN COMPROMISED, BUT SOME OF THE BOTTLES FLAGGED POSITIVE. CUSTOMER INDICATED THAT THEY FOUND NO INOCULATED BOTTLES WITH CRACKS PRIOR TO LOADING NOR DURING AN INSPECTION OF UNINOCULATED BOTTLES. THERE WAS NO PT OR RESULTS IMPACT DUE TO THIS ISSUE NOR WAS THERE ANY DEATHS OR INJURIES RELATED TO THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BACT/ALERT MP CULTURE BOTTLES | MICROBIAL GROWTH MONTIOR | MDB | BIOMERIEUX, INC. | 259797 | 1015498,1015712 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |