FDA Adverse Event Malfunction Summary report: N

SUMMIT SWIVEL COLLIMATOR

MDR report key: 871844 · Received June 26, 2007

Report

Report Number
1450503-2007-00002
Event Type
Malfunction
Date Received
June 26, 2007
Date of Event
June 8, 2007
Report Date
June 8, 2007
Manufacturer
SUMMIT INDUSTRIES, INC.
Product Code
IZX
PMA / PMN Number
K874687
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS INCIDENT IS RELATED TO MFR REPORT# 1450503-2007-00001.

Description of Event or Problem · 1

END USER CALLED SUMMIT DEALER TO ADDRESS "CUT-OFF" ON IMAGES. DEALER SVC REP WENT TO END USER FACILITY AND FOUND COLLIMATOR HANGING BY 2 SCREWS INSTEAD OF 4, TOLD SITE TO DISCONTINUE USE. SUBJECT COLLIMATOR SWIVEL MOUNTING ASSEMBLY WAS REPLACED AND RETURNED TO MANUFACTURER, SUMMIT INDUSTRIES, INC. FOR EVALUATION. THIS INCIDENT IS RELATED TO MFR REPORT# 1450503-2007-00001.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUMMIT SWIVEL COLLIMATOR BEAM LIMITING DEVICE IZX SUMMIT INDUSTRIES, INC. D800 NA851-1193

Patients

Seq Age Sex Outcome Treatment
1 YR