I-STAT 6+ CARTRIDGE
Report
- Report Number
- 2245578-2019-00155
- Event Type
- Malfunction
- Date Received
- June 20, 2019
- Date of Event
- April 20, 2019
- Report Date
- July 22, 2019
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- UDI-DI
- 10054749000088
- PMA / PMN Number
- K912387
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 07/19/2019. A REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THE LOTS PASSED FINISHED GOODS RELEASE CRITERIA. RETAIN CARTRIDGE TEST RESULTS FROM LOT N19105 MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AD, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. DUE TO THE EXPIRY DATE OF LOT K18209 (24-APR-19) RETAIN TESTING COULD NOT BE PERFORMED. RETAIN CARTRIDGE TEST RESULTS FROM A PREVIOUS INCIDENT FOR LOT K18209 DID NOT MEET THE ACCEPTANCE CRITERION FOR SUPPRESSED RESULTS FOUND IN Q04.01.003 REV. AD, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATIO,N PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED LOT N19105. A PREVIOUS UNRELATED DEFICIENCY WAS IDENTIFIED FOR HIGHER THAN EXPECTED RATE OF SUPPRESSED RESULTS FOR LOT K18209. QUALITY RECORD 681812 WAS INITIATED TO INVESTIGATE TO ROOT CAUSE.
APOC INCIDENT # (B)(4). OTHER LOT USED: PRODUCT#: 03P76-25, DESCRIPTION: EG7+, LOT#: N19105, EXPIRATION DATE: 02/13/2019, MFG DATE: 04/15/2019. APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON (B)(6) 2019, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT 6+ AND EG7+ CARTRIDGES THAT YIELDED A SUSPECTED A DISCREPANT SODIUM, POTASSIUM, HEMATOCRIT AND HEMOGLOBIN RESULT ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. (B)(6). THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE TESTING BETWEEN EG7+ AND 6+ CARTRIDGES WAS 1 HOUR 2 MINS. THE CUSTOMER WAS ADVISED THAT THE TEST NEEDS TO BE DONE WITHIN 30 MINUTES AFTER COLLECTION FOR NA, K, HCT & HB PER I-STAT SYSTEM MANUAL 714258-01 REV O. IT IS SUSPECTED THAT THE SECOND SAMPLE WAS CLOTTED, AND CAUSE FOR THE CUSTOMER COMPLAINT WAS DUE TO POOR SAMPLE HANDING. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 509137 | I-STAT 6+ CARTRIDGE | 6+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | K18209 | 10054749000088 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |