BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE
Report
- Report Number
- 3003152976-2019-00405
- Event Type
- Malfunction
- Date Received
- June 19, 2019
- Date of Event
- May 17, 2019
- Report Date
- September 30, 2019
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE COMPLAINT IS NOT REPORTABLE UPON FURTHER REVIEW. THE SAMPLES RETURNED SHOWED THAT FOREIGN MATTER WAS VISIBLE ON THE OUTSIDE OF THE PRODUCT. THIS IS NOT MDR REPORTABLE. THERE WAS NO REPORT OF SERIOUS INJURY, MEDICAL INTERVENTION, OR REPORTABLE DEVICE MALFUNCTION. FOREIGN MATTER NOT WITHIN THE FLUID PATHWAY MAY CAUSE A CUSTOMER INCONVENIENCE BUT WILL NOT LEAD TO ANY DEGREE OF HARM/SERIOUS INJURY. THE REPORT#: 3003152976-2019-00405 IS NULL AND VOID AS IT IS NOT LONGER MDR REPORTABLE.
H.6. INVESTIGATION SUMMARY: TEN PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. IN ALL PHOTOS, BLACK SPOTS CAN BE OBSERVED ON THE PLUNGER NERVES OR WITHIN THE BLISTER PACK. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOTS 1902257,1901240, AND1903215, NO DEVIATIONS OR NON-CONFORMANCES RELATED TO THE REPORTED ISSUE WERE IDENTIFIED DURING THE MANUFACTURING PROCESS. WITHOUT A PHYSICAL SAMPLE WE COULD NOT CONDUCT CHARACTERIZATION ANALYSIS ON THE FOREIGN MATTER TO FURTHER IDENTIFY THE ORIGIN. BASED ON THE PHOTO AND AVAILABLE INFORMATION, WE ARE NOT ABLE TO DETERMINED A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED ISSUE WILL CONTINUE TO BE TRACKED AND TRENDED FOR FUTURE OCCURRENCE. H3 OTHER TEXT : SEE SECTION H.10.
IT WAS REPORTED THAT BLACK PARTICLES WERE FOUND IN THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE BEFORE USE. LOT#'S 1902257, 1901240, AND 1903215 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: HERE ARE SOME REPRESENTATIVE PHOTOS OF THE BLACK PARTICLES OBSERVED ON LOTS: 1902257, 1901240, 1903215 SYRINGES OF 60ML.
IT WAS REPORTED THAT BLACK PARTICLES WERE FOUND IN THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE BEFORE USE. LOT #'S 1902257, 1901240, AND 1903215 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: HERE ARE SOME REPRESENTATIVE PHOTOS OF THE BLACK PARTICLES OBSERVED ON LOTS 1902257, 1901240, 1903215 SYRINGES OF 60ML.
THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1902257, MEDICAL DEVICE EXPIRATION DATE: 2024-01-31, DEVICE MANUFACTURE DATE: 2019-02-25. MEDICAL DEVICE LOT #: 1901240, MEDICAL DEVICE EXPIRATION DATE: 2023-12-31, DEVICE MANUFACTURE DATE: 2019-01-24. MEDICAL DEVICE LOT #: 1903215, MEDICAL DEVICE EXPIRATION DATE: 2024-02-29, DEVICE MANUFACTURE DATE:2019-03-21. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BLACK PARTICLES WERE FOUND IN THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE BEFORE USE. LOT #'S 1902257, 1901240, AND 1903215 WERE REPORTED TO HAVE BEEN INVOLVED IN THIS EVENT, BUT IT IS UNKNOWN HOW MANY OCCURRENCES HAPPENED WITHIN EACH LOT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: HERE ARE SOME REPRESENTATIVE PHOTOS OF THE BLACK PARTICLES OBSERVED ON LOTS 1902257, 1901240, 1903215 SYRINGES OF 60ML.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 508397 | BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | SEE SECTION H.10. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |